SyringeabilityForceTester
Syringeability Force Tester
Syringeability Force Tester — Filler Extrusion Force Test (ISO 13485)
ℹ️ The utility checks peak and average force, smoothness (CV), ergonomics.
ℹ️ Does not use machine learning — mechanical calculations only.
⚠️ CRITICAL! Force must match needle gauge. CV < 15% (smoothness).
Usage:
SyringeabilityForceTester.exe → demo mode
SyringeabilityForceTester.exe input.csv output.json → evaluate data
Input format:
BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type
Example:
SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall
— WHY IS THIS NEEDED?
Extrusion force determines physician comfort and injection accuracy:
• Peak Force (Max Force) — force to break static friction ("breakthrough"). Too high causes needle jump and tissue trauma.
• Average Force (Avg Force) — force to maintain flow. High force causes hand fatigue during large volume treatments.
• Smoothness (CV%) — coefficient of variation of force. High CV means "stick-slip" effect (jerky motion), making smooth lines/boluses impossible.
• Needle dependence: force increases sharply with decreasing gauge (27G -> 30G) and needle length.
⚠️ CRITICAL:
• Force ≤ Specification for given gauge (e.g., ≤15N for 27G, ≤25N for 30G).
• CV ≤ 15% — guarantee of smooth extrusion without jerks.
• Ergonomic Rating: assesses combination of force and smoothness (Excellent/Good/Poor).
• Rheology Correlation: high viscosity/G' should yield predictably high force. Discrepancies may indicate needle clogging or syringe defect.
• Sterility and Endotoxins — mandatory.
⚠️ Note: Utility accounts for needle type (ThinWall has lower resistance than Regular). Smoothness analysis (CV) often more important than absolute force, as jerks during facial injection are unacceptable. Test performed on texture analyzer or universal testing machine.input.csv
BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall
URS & FS — user requirements and functional specification
This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.
1. Scope
The Syringeability Force Tester utility is used for Syringeability Force Tester. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.
The utility does not use machine learning; decisions are produced by deterministic rules.
Categories: Dermal fillers
Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility
2. Execution modes
SyringeabilityForceTester.exe → demo mode SyringeabilityForceTester.exe input.csv output.json → evaluate data
3. Key controlled areas
- microbiology and pathogens
- pH and physicochemical parameters
4. Domain limits and critical parameters
- ⚠️ CRITICAL! Force must match needle gauge. CV < 15% (smoothness).
- Needle dependence: force increases sharply with decreasing gauge (27G -> 30G) and needle length.
- ⚠️ CRITICAL:
- Force ≤ Specification for given gauge (e.g., ≤15N for 27G, ≤25N for 30G).
- CV ≤ 15% — guarantee of smooth extrusion without jerks.
- Ergonomic Rating: assesses combination of force and smoothness (Excellent/Good/Poor).
- Rheology Correlation: high viscosity/G' should yield predictably high force. Discrepancies may indicate needle clogging or syringe defect.
- Sterility and Endotoxins — mandatory.
- ⚠️ Note: Utility accounts for needle type (ThinWall has lower resistance than Regular). Smoothness analysis (CV) often more important than absolute force, as jerks during facial injection are unacceptable. Test performed on texture analyzer or universal testing machine.
5. URS — user requirements
| ID | Requirement | Criticality | Acceptance criterion |
|---|---|---|---|
| URS-001 | The system shall accept an input.csv file for Syringeability Force Tester with the exact columns listed in the “Input data contract” section. | High | A file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error. |
| URS-002 | The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data. | Medium | Both execution scenarios produce a predictable result or clear diagnostic error. |
| URS-003 | The system shall perform rule-based controls for: microbiology and pathogens, pH and physicochemical parameters. | High | Each controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML. |
| URS-004 | The system shall preserve traceability between batch/lot, source values, applied rules and final verdict. | High | Output includes batch identifier, source values, parameter statuses and critical findings. |
| URS-005 | The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review. | High | JSON contains overall status, check array, warnings, failures and source-file reference. |
| URS-006 | The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks. | High | The document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval. |
| URS-007 | The system shall clearly separate technical data errors from specification nonconformities. | Medium | Schema/type errors are not mixed with pharmacopoeial deviations and are reported separately. |
6. input.csv data contract
The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.
| # | Column | Type | Unit | Description | Sample | Control rule |
|---|---|---|---|---|---|---|
| 1 | BatchNumber | identifier | as specified | Batch Number | SYR-HA-2026-001 | mandatory field; used for batch/lot traceability |
| 2 | Max_Force_N | decimal | as specified | Max Force N | 12.5 | mandatory numeric value; rule comparison is performed by the utility |
| 3 | Average_Force_N | decimal | as specified | Average Force N | 9.0 | mandatory numeric value; rule comparison is performed by the utility |
| 4 | Syringe_Barrel_Diameter_mm | decimal | as specified | Syringe Barrel Diameter mm | 9.5 | mandatory numeric value; rule comparison is performed by the utility |
| 5 | Needle_Gauge | decimal | as specified | Needle Gauge | 27.0 | mandatory numeric value; rule comparison is performed by the utility |
| 6 | Needle_Length_mm | decimal | as specified | Needle Length mm | 13.0 | mandatory numeric value; rule comparison is performed by the utility |
| 7 | Plunger_Speed_mm_per_min | decimal | min | Plunger Speed mm per min | 30.0 | mandatory numeric value; rule comparison is performed by the utility |
| 8 | Temperature_C | decimal | °C | Temperature C | 25.0 | mandatory numeric value; rule comparison is performed by the utility |
| 9 | Viscosity_Pa_s | decimal | as specified | Viscosity Pa s | 25.0 | mandatory numeric value; rule comparison is performed by the utility |
| 10 | G_Prime_Pa | decimal | as specified | G Prime Pa | 450.0 | mandatory numeric value; rule comparison is performed by the utility |
| 11 | Max_Allowable_Force_N | decimal | as specified | Max Allowable Force N | 15.0 | mandatory numeric value; rule comparison is performed by the utility |
| 12 | Force_Variation_CV_Percent | decimal | % | Force Variation CV % | 8.5 | mandatory numeric value; rule comparison is performed by the utility |
| 13 | SterilityTest | boolean / flag | 0/1 | Sterility Test | 0 | valid flag required; prohibited organisms and growth are normally expected as 0 / absent |
| 14 | Endotoxin_EU_ml | decimal | as specified | Endotoxin EU ml | 0.1 | numeric value; critical safety attribute compared with endotoxin limit |
| 15 | Needle_Type | text | as specified | Needle Type | ThinWall | mandatory value; format and acceptability are checked by the utility |
CSV example
BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall
7. FS — functional specification
| ID | Function | Implementation description |
|---|---|---|
| FS-001 | CLI entry point | The executable SyringeabilityForceTester.exe supports demo mode and input.csv output.json processing mode. |
| FS-002 | CSV parser | The import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator. |
| FS-003 | Field conversion | Each column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag. |
| FS-004 | Domain rule engine | For Syringeability Force Tester, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check. |
| FS-005 | Criticality handling | Critical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS. |
| FS-006 | JSON writer | output.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages. |
| FS-007 | Integration contract | The CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service. |
| FS-008 | Error handling | Schema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS. |
8. output.json contract
The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.
{
"utility": "SyringeabilityForceTester",
"api": "Syringeability Force Tester",
"batchNumber": "SYR-HA-2026-001",
"overallStatus": "PASS|WARNING|FAIL",
"checkedAtUtc": "2026-05-18T00:00:00Z",
"checks": [
{
"parameter": "BatchNumber",
"value": "SYR-HA-2026-001",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Max_Force_N",
"value": "12.5",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Average_Force_N",
"value": "9.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Syringe_Barrel_Diameter_mm",
"value": "9.5",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Needle_Gauge",
"value": "27.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Needle_Length_mm",
"value": "13.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Plunger_Speed_mm_per_min",
"value": "30.0",
"unit": "min",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Temperature_C",
"value": "25.0",
"unit": "°C",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
}
],
"criticalFindings": [],
"sourceFile": "input.csv"
}9. OQ/PQ test scenarios
| ID | Scenario | Expected result |
|---|---|---|
| TC-001 | Valid CSV with expected header and sample row | All rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail. |
| TC-002 | A mandatory input.csv column is missing | Import is rejected or the row receives FAIL with schema reference. |
| TC-003 | A numeric field contains text or a blank value | Type conversion error is recorded; the result is not hidden as PASS. |
| TC-004 | A parameter is outside specification or critical limit | Critical parameters produce FAIL; non-critical deviations produce WARNING according to the rule. |
| TC-005 | Positive pathogen/microbiological flag or microbiological limit excursion | Critical FAIL is produced with the offending parameter. |
10. QA/QC, CSV and change control
- Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
- Column names must not be changed without updating the validator and test scenarios.
- The source CSV, output.json and utility version should be stored together as an evidence package.
- Any change in control rules must go through change control and repeated OQ scenario verification.
Included in packages
Dermal Fillers QC Suite
Package for injectable implants and dermal fillers: hyaluronic acid, CaHA, PLLA, PMMA, collagen and related tests.
Open