SmallFieldCorrectionChecker
Small Field Correction
Small Field Correction Checker — Small Field Correction Verification (IAEA TRS-483)
ℹ️ The utility verifies correctness of applied detector correction factors ($k_{Q_{clin},Q_{msr}}^{f_{clin},f_{msr}}$) for small fields (SRS, SBRT) by comparing with reference data (IAEA TRS-483).
ℹ️ Critical for fields < 2×2 cm where detector volume averaging and density effects cause significant measurement errors.
ℹ️ Processing: only the FIRST record in input.csv file.
⚠️ Critical for safety! A 5% error in correction factor for 1 cm field leads to direct 5% dose error to patient, unacceptable in stereotaxy.
Usage:
SmallFieldCorrectionChecker.exe → demo mode
SmallFieldCorrectionChecker.exe input.csv output.json → check data (first record only)
Input format:
PatientID,FieldName,FieldSizeCm,DetectorType,MeasuredOutputFactor,AppliedCorrectionFactor,ReferenceCorrectionFactor
Example (1.0 cm field, diode detector):
PT-2026-BRAIN-SRS,SRS_Cone_10mm,1.0,Diode_E,0.780,0.920,0.965
— WHY IS THIS NEEDED?
In small field dosimetry, standard detectors underestimate signal due to volume averaging and material density.
• Correction factors from protocols (TRS-483) must be applied.
• Physicist might err by selecting factor for wrong field size or detector type.
• Utility automatically compares entered value with reference and identifies discrepancies.
• Allows catching error before plan approval.
⚠️ CRITICAL:
• Tolerance on deviation: ≤ 1.5% (Warning), > 3.0% (Fail).
• For fields < 1 cm requirements are even stricter.
• Cannot use factor 1.0 for fields < 2 cm (unless special detector).
• TRS-483 tables must be up-to-date.
Critical parameters:
• Field size: < 2.0 cm (correction zone)
• Factor deviation: > 3% — critical error
• Detector type: Diodes, Diamonds, Ion chambers require different corrections
⚠️ Note: Utility flags cases where factor 1.0 is used for small field. Result includes status (PASS/WARNING/FAIL) and bilingual recommendation (RU/EN).input.csv
PatientID,FieldName,FieldSizeCm,DetectorType,MeasuredOutputFactor,AppliedCorrectionFactor,ReferenceCorrectionFactor PT-2026-BRAIN-SRS,SRS_Cone_10mm,1.0,Diode_E,0.780,0.920,0.965
URS & FS — user requirements and functional specification
This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.
1. Scope
The Small Field Correction Checker utility is used for Small Field Correction. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.
The utility does not use machine learning; decisions are produced by deterministic rules.
Categories: Radiation physics
Tags: activity, brachytherapy, dosimetry, source calibration, TG-43
2. Execution modes
SmallFieldCorrectionChecker.exe → demo mode SmallFieldCorrectionChecker.exe input.csv output.json → check data (first record only)
3. Key controlled areas
- pH and physicochemical parameters
- activity, decay and radiochemical purity
4. Domain limits and critical parameters
- ⚠️ Critical for safety! A 5% error in correction factor for 1 cm field leads to direct 5% dose error to patient, unacceptable in stereotaxy.
- ⚠️ CRITICAL:
- Tolerance on deviation: ≤ 1.5% (Warning), > 3.0% (Fail).
- For fields < 1 cm requirements are even stricter.
- Cannot use factor 1.0 for fields < 2 cm (unless special detector).
- TRS-483 tables must be up-to-date.
- Critical parameters:
- Field size: < 2.0 cm (correction zone)
- Factor deviation: > 3% — critical error
- Detector type: Diodes, Diamonds, Ion chambers require different corrections
- ⚠️ Note: Utility flags cases where factor 1.0 is used for small field. Result includes status (PASS/WARNING/FAIL) and bilingual recommendation (RU/EN).
5. URS — user requirements
| ID | Requirement | Criticality | Acceptance criterion |
|---|---|---|---|
| URS-001 | The system shall accept an input.csv file for Small Field Correction with the exact columns listed in the “Input data contract” section. | High | A file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error. |
| URS-002 | The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data. | Medium | Both execution scenarios produce a predictable result or clear diagnostic error. |
| URS-003 | The system shall perform rule-based controls for: pH and physicochemical parameters, activity, decay and radiochemical purity. | High | Each controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML. |
| URS-004 | The system shall preserve traceability between batch/lot, source values, applied rules and final verdict. | High | Output includes batch identifier, source values, parameter statuses and critical findings. |
| URS-005 | The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review. | High | JSON contains overall status, check array, warnings, failures and source-file reference. |
| URS-006 | The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks. | High | The document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval. |
| URS-007 | The system shall clearly separate technical data errors from specification nonconformities. | Medium | Schema/type errors are not mixed with pharmacopoeial deviations and are reported separately. |
6. input.csv data contract
The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.
| # | Column | Type | Unit | Description | Sample | Control rule |
|---|---|---|---|---|---|---|
| 1 | PatientID | text | as specified | Patient ID | PT-2026-BRAIN-SRS | mandatory value; format and acceptability are checked by the utility |
| 2 | FieldName | text | as specified | Field Name | SRS_Cone_10mm | mandatory value; format and acceptability are checked by the utility |
| 3 | FieldSizeCm | decimal | as specified | Field Size Cm | 1.0 | mandatory numeric value; rule comparison is performed by the utility |
| 4 | DetectorType | text | as specified | Detector Type | Diode_E | mandatory value; format and acceptability are checked by the utility |
| 5 | MeasuredOutputFactor | decimal | as specified | Measured Output Factor | 0.780 | mandatory numeric value; rule comparison is performed by the utility |
| 6 | AppliedCorrectionFactor | decimal | as specified | Applied Correction Factor | 0.920 | mandatory numeric value; rule comparison is performed by the utility |
| 7 | ReferenceCorrectionFactor | decimal | as specified | Reference Correction Factor | 0.965 | mandatory numeric value; rule comparison is performed by the utility |
CSV example
PatientID,FieldName,FieldSizeCm,DetectorType,MeasuredOutputFactor,AppliedCorrectionFactor,ReferenceCorrectionFactor PT-2026-BRAIN-SRS,SRS_Cone_10mm,1.0,Diode_E,0.780,0.920,0.965
7. FS — functional specification
| ID | Function | Implementation description |
|---|---|---|
| FS-001 | CLI entry point | The executable SmallFieldCorrectionChecker.exe supports demo mode and input.csv output.json processing mode. |
| FS-002 | CSV parser | The import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator. |
| FS-003 | Field conversion | Each column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag. |
| FS-004 | Domain rule engine | For Small Field Correction, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check. |
| FS-005 | Criticality handling | Critical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS. |
| FS-006 | JSON writer | output.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages. |
| FS-007 | Integration contract | The CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service. |
| FS-008 | Error handling | Schema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS. |
8. output.json contract
The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.
{
"utility": "SmallFieldCorrectionChecker",
"api": "Small Field Correction",
"batchNumber": "",
"overallStatus": "PASS|WARNING|FAIL",
"checkedAtUtc": "2026-05-18T00:00:00Z",
"checks": [
{
"parameter": "PatientID",
"value": "PT-2026-BRAIN-SRS",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "FieldName",
"value": "SRS_Cone_10mm",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "FieldSizeCm",
"value": "1.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "DetectorType",
"value": "Diode_E",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "MeasuredOutputFactor",
"value": "0.780",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "AppliedCorrectionFactor",
"value": "0.920",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "ReferenceCorrectionFactor",
"value": "0.965",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
}
],
"criticalFindings": [],
"sourceFile": "input.csv"
}9. OQ/PQ test scenarios
| ID | Scenario | Expected result |
|---|---|---|
| TC-001 | Valid CSV with expected header and sample row | All rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail. |
| TC-002 | A mandatory input.csv column is missing | Import is rejected or the row receives FAIL with schema reference. |
| TC-003 | A numeric field contains text or a blank value | Type conversion error is recorded; the result is not hidden as PASS. |
| TC-004 | A parameter is outside specification or critical limit | Critical parameters produce FAIL; non-critical deviations produce WARNING according to the rule. |
10. QA/QC, CSV and change control
- Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
- Column names must not be changed without updating the validator and test scenarios.
- The source CSV, output.json and utility version should be stored together as an evidence package.
- Any change in control rules must go through change control and repeated OQ scenario verification.
Included in packages
Radiation Physics QC Suite
Utilities for activity, dosimetry, brachytherapy sources, small-field corrections and TG-43 parameters.
OpenRPH organ/system: brain and CNS
Brain perfusion, dopamine transporter, amyloid/tau, receptor imaging and neurological SPECT/PET.
OpenRPH use case: PET scanner state, gamma camera and radiation physics
The linked medical-device/radiation-physics layer: PET detector QC, normalization, AC map, cross-calibration, TOF, gamma-camera uniformity, source strength and staff dose.
OpenRPH workflow: dosimetry and therapy planning