Sequencer Run QC Checker — Comprehensive NGS Sequencer Run Quality Control

Performs instrument-level run QC per CAP/CLIA §GEN.48800, ISO 15189 §7.3.4, FDA NGS guidance and manufacturer specifications: clustering/loading metrics (cluster density within range, PF% ≥75%, phasing/prephasing ≤1.0%, occupancy ≥70% for patterned FCs), base quality (Q30 R1/R2 ≥75%, Q40 for newer chemistries, quality drop cycle detection), output metrics (yield ≥80% expected, alignment rate ≥70%), controls (PhiX error rate ≤2.0%, spike-in passed, negative control clean), flow cell uniformity (lane CV ≤15%, tile hotspots, index hopping ≤0.5%), instrument health (calibration current, maintenance up-to-date, error codes, fluidics/optics issues, temperature stability CV ≤0.5), and consumables verification (lot tracking, expiry check). Three-tier classification: PASS / REVIEW_REQUIRED / FAIL.

⚠️ Critical: PhiX error rate is the most sensitive early indicator of sequencing problems. Overclustering irreversibly degrades quality; underloading preferred over overloading. Expired consumables = legally invalid data for clinical use. Fluidics issues manifest as high phasing + lane non-uniformity. Index hopping on patterned FCs without UDI can reach 5–10%. Quality drop at specific cycle indicates nucleotide or chemistry cycle issue. Temperature instability affects polymerase kinetics and reduces Q-scores. Tool assesses INSTRUMENT-LEVEL RUN QUALITY only; does not replace sample/library-level QC or bioinformatic variant validation. All three levels required. FAIL status data must not enter bioinformatics pipeline until issue resolved or run repeated.

Usage:
  SequencerRunQcChecker.exe                            → demo mode (console output)
  SequencerRunQcChecker.exe input.csv output.json      → evaluate your data

📍 Scope of Application:
     • NGS laboratories: Mandatory acceptance of every run before bioinformatics.
     • Core facilities: Standardized QC for research projects.
     • IVD manufacturing sites: Sequencer release testing.
     • Regulatory inspections: Documentation of instrument performance monitoring.
     • Field service engineers: Instrument diagnostics via run metrics.

💡 Tips:
1. Platform-Specific Thresholds: Configure separate threshold profiles for each platform/flow cell type.
2. Trend Monitoring: Track Q30, PF%, PhiX error rate over time for predictive maintenance.
3. Control Charts: Maintain Levey-Jennings for key metrics; trends matter more than single values.
4. Consumable Tracking: Integrate lot tracking with LIMS for automatic expiry verification.
5. Error Code Dictionary: Create error code reference with recommended corrective actions.

⚠️ Note: This utility assesses SEQUENCER INSTRUMENT-LEVEL RUN QUALITY. It does not replace sample/library-level QC (LibraryPrepBatchChecker, PanelCoverageQcChecker) or bioinformatic variant validation. All three levels are required: instrument QC → library/sample QC → bioinformatic QC. Data from FAIL runs must not enter the bioinformatics pipeline until the issue is resolved or the run is repeated.

input.csv

RunID,Platform,InstrumentSerial,FlowCell_Type,Chemistry_Version,ReadConfiguration,RunStartDate,RunEndDate,RunDuration_Hours,ClusterDensity_K_mm2,MinClusterDensity_K_mm2,MaxClusterDensity_K_mm2,PercentPF,MinPercentPF,PhasingRate_Percent,MaxPhasingRate_Percent,PrephasingRate_Percent,MaxPrephasingRate_Percent,Occupancy_Percent,MinOccupancy_Percent,PercentQ30_Read1,PercentQ30_Read2,PercentQ30_Index1,PercentQ30_Index2,MinPercentQ30,PercentQ40_Read1,MedianQualityScore,QualityDrop_Cycle,TotalReads_Million,ExpectedReads_Million,Yield_Gb,ExpectedYield_Gb,PercentAligned,MinPercentAligned,PhiX_ErrorRate_Percent,MaxPhiX_ErrorRate_Percent,PhiX_AlignmentRate_Percent,SpikeIn_Control_Passed,NegativeControl_Clean,LaneUniformity_CV_Percent,MaxLaneUniformity_CV_Percent,TileHotspots_Detected,IndexHopping_Rate_Percent,MaxIndexHopping_Rate_Percent,Calibration_Current,Maintenance_UpToDate,ErrorCodes_DuringRun,FluidicsIssues_Detected,OpticsIssues_Detected,TemperatureStability_CV,MaxTemperatureStability_CV,ReagentKit_Lot,FlowCell_Lot,Consumables_Expired,ReagentKit_RemainingCycles
NS6000-2026-0142,NovaSeq_6000,A01234,S4,v1.5,2x150,2026-06-07T08:00:00,2026-06-08T04:00:00,20,950,600,1200,88.5,75,0.15,1.0,0.12,1.0,92.0,70,92.5,89.8,95.2,94.8,75,78.5,36,0,2800,3000,840,900,96.5,70,0.18,2.0,98.5,true,true,4.2,15,false,0.03,0.5,true,true,,false,false,0.08,0.5,RK-2026-0451,FC-2026-0892,false,0
NS6000-2026-0143,NovaSeq_6000,A01234,SP,v1.5,2x100,2026-06-08T10:00:00,2026-06-08T22:00:00,12,1350,600,1200,58.0,75,2.8,1.0,1.9,1.0,95.0,70,68.5,55.2,82.0,78.5,75,35.0,28,45,450,800,90,160,72.0,70,3.5,2.0,92.0,false,true,22.5,15,true,1.8,0.5,true,false,"ERR_FLUIDICS_BUBBLE;ERR_TEMP_SPIKE",true,false,0.85,0.5,RK-2026-0467,FC-2026-0901,false,0

URS & FS — User Requirements and Functional Specification

This document describes the controlled interface and behaviour of SequencerRunQcChecker for Sequencer Run QC Checker.

Domain limits and critical parameters

Key fragments from the source description are shown below. Before production use, limits must be verified against the approved specification, registration dossier and local SOPs.
  • Performs instrument-level run QC per CAP/CLIA §GEN.48800, ISO 15189 §7.3.4, FDA NGS guidance and manufacturer specifications: clustering/loading metrics (cluster density within range, PF% ≥75%, phasing/prephasing ≤1.0%, occupancy ≥70% for patterned FCs), base quality (Q30 R1/R2 ≥75%, Q40 for newer chemistries, quality drop cycle detection), output metrics (yield ≥80% expected, alignment rate ≥70%), controls (PhiX error rate ≤2.0%, spike-in passed, negative control clean), flow cell uniformity (l
  • ⚠️ Critical: PhiX error rate is the most sensitive early indicator of sequencing problems. Overclustering irreversibly degrades quality; underloading preferred over overloading. Expired consumables = legally invalid data for clinical use. Fluidics issues manifest as high phasing + lane non-uniformity. Index hopping on patterned FCs without UDI can reach 5–10%. Quality drop at specific cycle indicates nucleotide or chemistry cycle issue. Temperature instability affects polymerase kinetics and red
  • ⚠️ Note: This utility assesses SEQUENCER INSTRUMENT-LEVEL RUN QUALITY. It does not replace sample/library-level QC (LibraryPrepBatchChecker, PanelCoverageQcChecker) or bioinformatic variant validation. All three levels are required: instrument QC → library/sample QC → bioinformatic QC. Data from FAIL runs must not enter the bioinformatics pipeline until the issue is resolved or the run is repeated.

URS — User Requirements Specification

IDRequirementCriticalityAcceptance criterion
URS-001The utility shall accept an input.csv file for Sequencer Run QC Checker with headers defined in the data contract.HighThe file is processed without manual header editing.
URS-002The utility shall perform deterministic QC evaluation without machine learning and without probabilistic conformance decisions.HighIdentical input data, rule version and configuration produce reproducible results.
URS-003The utility shall validate mandatory fields, data types, ranges, units and domain plausibility.HighSchema, conversion and range errors are explicitly reported.
URS-004The utility shall apply domain limits and rules from the description, approved specification, registration dossier and local SOPs.HighEach check has PASS/WARNING/FAIL and a clear message.
URS-005The utility shall generate output.json with machine-readable results, source values, warnings, failures and critical findings.HighJSON is suitable for LIMS/ELN/MES integration and QA/QC review.
URS-006The utility shall preserve traceability between batch/sample, input file, applied rules and final status.HighOutput contains identifiers, checked parameters and audit metadata.
URS-007The documentation shall support IQ/OQ/PQ, CSV/CSA and review by internal QA or inspectors.MediumURS, FS, input/output contract and test scenarios are supplied with the utility.
URS-008The utility shall be used as a QC decision-support tool and not as a substitute for approved specifications and QA/QP release decision.MediumDocumentation states change control and limit-verification expectations.

input.csv contract

#FieldTypeSamplePurpose
1RunIDstring / controlled vocabularyNS6000-2026-0142Controlled input parameter for deterministic QC rules.
2Platformstring / controlled vocabularyNovaSeq_6000Controlled input parameter for deterministic QC rules.
3InstrumentSerialstring / controlled vocabularyA01234Controlled input parameter for deterministic QC rules.
4FlowCell_Typestring / controlled vocabularyS4Controlled input parameter for deterministic QC rules.
5Chemistry_Versionstring / controlled vocabularyv1.5Controlled input parameter for deterministic QC rules.
6ReadConfigurationstring / controlled vocabulary2x150Component ratio; structural or formulation CQA.
7RunStartDatestring / controlled vocabulary2026-06-07T08:00:00Controlled input parameter for deterministic QC rules.
8RunEndDatestring / controlled vocabulary2026-06-08T04:00:00Controlled input parameter for deterministic QC rules.
9RunDuration_Hoursinteger / decimal20Time parameter for process, incubation, storage or analytical workflow.
10ClusterDensity_K_mm2decimal950Controlled input parameter for deterministic QC rules.
11MinClusterDensity_K_mm2decimal600Controlled input parameter for deterministic QC rules.
12MaxClusterDensity_K_mm2decimal1200Controlled input parameter for deterministic QC rules.
13PercentPFdecimal88.5Controlled input parameter for deterministic QC rules.
14MinPercentPFdecimal75Controlled input parameter for deterministic QC rules.
15PhasingRate_Percentdecimal0.15pH; parameter relevant to stability, compatibility and batch suitability.
16MaxPhasingRate_Percentdecimal1.0pH; parameter relevant to stability, compatibility and batch suitability.
17PrephasingRate_Percentdecimal0.12pH; parameter relevant to stability, compatibility and batch suitability.
18MaxPrephasingRate_Percentdecimal1.0pH; parameter relevant to stability, compatibility and batch suitability.
19Occupancy_Percentdecimal92.0Controlled input parameter for deterministic QC rules.
20MinOccupancy_Percentdecimal70Controlled input parameter for deterministic QC rules.
21PercentQ30_Read1decimal92.5Controlled input parameter for deterministic QC rules.
22PercentQ30_Read2decimal89.8Controlled input parameter for deterministic QC rules.
23PercentQ30_Index1decimal95.2Controlled input parameter for deterministic QC rules.
24PercentQ30_Index2decimal94.8Controlled input parameter for deterministic QC rules.
25MinPercentQ30decimal75Controlled input parameter for deterministic QC rules.
26PercentQ40_Read1decimal78.5Controlled input parameter for deterministic QC rules.
27MedianQualityScoredecimal36Controlled input parameter for deterministic QC rules.
28QualityDrop_Cycledecimal0Controlled input parameter for deterministic QC rules.
29TotalReads_Millioninteger / decimal2800Controlled input parameter for deterministic QC rules.
30ExpectedReads_Millioninteger / decimal3000Controlled input parameter for deterministic QC rules.
31Yield_Gbdecimal840Controlled input parameter for deterministic QC rules.
32ExpectedYield_Gbdecimal900Controlled input parameter for deterministic QC rules.
33PercentAligneddecimal96.5Controlled input parameter for deterministic QC rules.
34MinPercentAligneddecimal70Controlled input parameter for deterministic QC rules.
35PhiX_ErrorRate_Percentdecimal0.18pH; parameter relevant to stability, compatibility and batch suitability.
36MaxPhiX_ErrorRate_Percentdecimal2.0pH; parameter relevant to stability, compatibility and batch suitability.
37PhiX_AlignmentRate_Percentdecimal98.5pH; parameter relevant to stability, compatibility and batch suitability.
38SpikeIn_Control_Passedstring / controlled vocabularytrueControl point/control sample for assay run validity.
39NegativeControl_Cleanstring / controlled vocabularytrueControl point/control sample for assay run validity.
40LaneUniformity_CV_Percentdecimal4.2Controlled input parameter for deterministic QC rules.
41MaxLaneUniformity_CV_Percentdecimal15Controlled input parameter for deterministic QC rules.
42TileHotspots_Detectedstring / controlled vocabularyfalseControlled input parameter for deterministic QC rules.
43IndexHopping_Rate_Percentdecimal0.03Controlled input parameter for deterministic QC rules.
44MaxIndexHopping_Rate_Percentdecimal0.5Controlled input parameter for deterministic QC rules.
45Calibration_Currentstring / controlled vocabularytrueComponent ratio; structural or formulation CQA.
46Maintenance_UpToDatestring / controlled vocabularytrueControlled input parameter for deterministic QC rules.
47ErrorCodes_DuringRunstring / controlled vocabularyControlled input parameter for deterministic QC rules.
48FluidicsIssues_Detectedstring / controlled vocabularyfalseControlled input parameter for deterministic QC rules.
49OpticsIssues_Detectedstring / controlled vocabularyfalseControlled input parameter for deterministic QC rules.
50TemperatureStability_CVdecimal0.08Temperature profile/MKT; storage, shipping and stability parameter.
51MaxTemperatureStability_CVdecimal0.5Temperature profile/MKT; storage, shipping and stability parameter.
52ReagentKit_Lotstring / controlled vocabularyRK-2026-0451Batch or lot identifier used for traceability.
53FlowCell_Lotstring / controlled vocabularyFC-2026-0892Batch or lot identifier used for traceability.
54Consumables_Expiredstring / controlled vocabularyfalseControlled input parameter for deterministic QC rules.
55ReagentKit_RemainingCyclesinteger / decimal0Controlled input parameter for deterministic QC rules.
RunID,Platform,InstrumentSerial,FlowCell_Type,Chemistry_Version,ReadConfiguration,RunStartDate,RunEndDate,RunDuration_Hours,ClusterDensity_K_mm2,MinClusterDensity_K_mm2,MaxClusterDensity_K_mm2,PercentPF,MinPercentPF,PhasingRate_Percent,MaxPhasingRate_Percent,PrephasingRate_Percent,MaxPrephasingRate_Percent,Occupancy_Percent,MinOccupancy_Percent,PercentQ30_Read1,PercentQ30_Read2,PercentQ30_Index1,PercentQ30_Index2,MinPercentQ30,PercentQ40_Read1,MedianQualityScore,QualityDrop_Cycle,TotalReads_Million,ExpectedReads_Million,Yield_Gb,ExpectedYield_Gb,PercentAligned,MinPercentAligned,PhiX_ErrorRate_Percent,MaxPhiX_ErrorRate_Percent,PhiX_AlignmentRate_Percent,SpikeIn_Control_Passed,NegativeControl_Clean,LaneUniformity_CV_Percent,MaxLaneUniformity_CV_Percent,TileHotspots_Detected,IndexHopping_Rate_Percent,MaxIndexHopping_Rate_Percent,Calibration_Current,Maintenance_UpToDate,ErrorCodes_DuringRun,FluidicsIssues_Detected,OpticsIssues_Detected,TemperatureStability_CV,MaxTemperatureStability_CV,ReagentKit_Lot,FlowCell_Lot,Consumables_Expired,ReagentKit_RemainingCycles
NS6000-2026-0142,NovaSeq_6000,A01234,S4,v1.5,2x150,2026-06-07T08:00:00,2026-06-08T04:00:00,20,950,600,1200,88.5,75,0.15,1.0,0.12,1.0,92.0,70,92.5,89.8,95.2,94.8,75,78.5,36,0,2800,3000,840,900,96.5,70,0.18,2.0,98.5,true,true,4.2,15,false,0.03,0.5,true,true,,false,false,0.08,0.5,RK-2026-0451,FC-2026-0892,false,0
NS6000-2026-0143,NovaSeq_6000,A01234,SP,v1.5,2x100,2026-06-08T10:00:00,2026-06-08T22:00:00,12,1350,600,1200,58.0,75,2.8,1.0,1.9,1.0,95.0,70,68.5,55.2,82.0,78.5,75,35.0,28,45,450,800,90,160,72.0,70,3.5,2.0,92.0,false,true,22.5,15,true,1.8,0.5,true,false,ERR_FLUIDICS_BUBBLE;ERR_TEMP_SPIKE,true,false,0.85,0.5,RK-2026-0467,FC-2026-0901,false,0

Input validation rules

IDFieldRuleCriticality
VR-001RunIDThe field shall match an approved dictionary or accepted string representation.High
VR-002PlatformThe field shall match an approved dictionary or accepted string representation.High
VR-003InstrumentSerialThe field shall match an approved dictionary or accepted string representation.High
VR-004FlowCell_TypeThe field shall match an approved dictionary or accepted string representation.Medium
VR-005Chemistry_VersionThe field shall match an approved dictionary or accepted string representation.Medium
VR-006ReadConfigurationThe field shall match an approved dictionary or accepted string representation.Medium
VR-007RunStartDateThe field shall match an approved dictionary or accepted string representation.Medium
VR-008RunEndDateThe field shall match an approved dictionary or accepted string representation.Medium
VR-009RunDuration_HoursThe field shall match an approved dictionary or accepted string representation.Medium
VR-010ClusterDensity_K_mm2The field shall match an approved dictionary or accepted string representation.Medium
VR-011MinClusterDensity_K_mm2The field shall match an approved dictionary or accepted string representation.Medium
VR-012MaxClusterDensity_K_mm2The field shall match an approved dictionary or accepted string representation.Medium
VR-013PercentPFThe field shall match an approved dictionary or accepted string representation.Medium
VR-014MinPercentPFThe field shall match an approved dictionary or accepted string representation.Medium
VR-015PhasingRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-016MaxPhasingRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-017PrephasingRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-018MaxPrephasingRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-019Occupancy_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-020MinOccupancy_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-021PercentQ30_Read1The field shall match an approved dictionary or accepted string representation.Medium
VR-022PercentQ30_Read2The field shall match an approved dictionary or accepted string representation.Medium
VR-023PercentQ30_Index1The field shall match an approved dictionary or accepted string representation.Medium
VR-024PercentQ30_Index2The field shall match an approved dictionary or accepted string representation.Medium
VR-025MinPercentQ30The field shall match an approved dictionary or accepted string representation.Medium
VR-026PercentQ40_Read1The field shall match an approved dictionary or accepted string representation.Medium
VR-027MedianQualityScoreThe field shall match an approved dictionary or accepted string representation.Medium
VR-028QualityDrop_CycleThe field shall match an approved dictionary or accepted string representation.Medium
VR-029TotalReads_MillionThe field shall match an approved dictionary or accepted string representation.Medium
VR-030ExpectedReads_MillionThe field shall match an approved dictionary or accepted string representation.Medium
VR-031Yield_GbThe field shall match an approved dictionary or accepted string representation.Medium
VR-032ExpectedYield_GbThe field shall match an approved dictionary or accepted string representation.Medium
VR-033PercentAlignedThe field shall match an approved dictionary or accepted string representation.Medium
VR-034MinPercentAlignedThe field shall match an approved dictionary or accepted string representation.Medium
VR-035PhiX_ErrorRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-036MaxPhiX_ErrorRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-037PhiX_AlignmentRate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-038SpikeIn_Control_PassedThe field shall match an approved dictionary or accepted string representation.Medium
VR-039NegativeControl_CleanThe field shall match an approved dictionary or accepted string representation.Medium
VR-040LaneUniformity_CV_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-041MaxLaneUniformity_CV_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-042TileHotspots_DetectedThe field shall match an approved dictionary or accepted string representation.Medium
VR-043IndexHopping_Rate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-044MaxIndexHopping_Rate_PercentThe field shall match an approved dictionary or accepted string representation.Medium
VR-045Calibration_CurrentThe field shall match an approved dictionary or accepted string representation.Medium
VR-046Maintenance_UpToDateThe field shall match an approved dictionary or accepted string representation.Medium
VR-047ErrorCodes_DuringRunThe field shall match an approved dictionary or accepted string representation.Medium
VR-048FluidicsIssues_DetectedThe field shall match an approved dictionary or accepted string representation.Medium
VR-049OpticsIssues_DetectedThe field shall match an approved dictionary or accepted string representation.Medium
VR-050TemperatureStability_CVThe field shall match an approved dictionary or accepted string representation.Medium
VR-051MaxTemperatureStability_CVThe field shall match an approved dictionary or accepted string representation.Medium
VR-052ReagentKit_LotThe field shall match an approved dictionary or accepted string representation.Medium
VR-053FlowCell_LotThe field shall match an approved dictionary or accepted string representation.Medium
VR-054Consumables_ExpiredThe field shall match an approved dictionary or accepted string representation.Medium
VR-055ReagentKit_RemainingCyclesThe field shall match an approved dictionary or accepted string representation.Medium

FS — Functional Specification

IDFunctionImplementation
FS-001CLI executionSupport execution modes: demo mode without arguments and production mode input.csv output.json.
FS-002CSV importRead input.csv in UTF-8/CSV-compatible format and validate header and expected columns.
FS-003Schema validationCheck mandatory fields, column count, unknown key fields and empty mandatory values.
FS-004Type conversionConvert numeric, flag and text values; invalid format is recorded as a row-level error.
FS-005Domain rule engineApply rules for Sequencer Run QC Checker, including critical limits from the description and approved specification.
FS-006Status aggregationProduce final status: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance.
FS-007JSON exportWrite output.json with detailed checks, source values, warnings, failures and critical findings.
FS-008Audit supportKeep result structure suitable for review, deviation investigation and calculation reproduction.
FS-009Integration contractSupport the scenario LIMS/ELN/MES → input.csv → utility → output.json → portal/admin review.
FS-010Error handlingReturn explicit messages for missing file, empty CSV, invalid schema, output write failure and invalid format.

Example output.json

{
  "utilityId": "sequencerrunqcchecker",
  "utilityFolder": "SequencerRunQcChecker",
  "package": "LiquidBiopsy",
  "overallStatus": "PASS|WARNING|FAIL",
  "sourceFile": "input.csv",
  "processedAtUtc": "2026-06-10T00:00:00Z",
  "checks": [
    {
      "parameter": "RunID",
      "value": "NS6000-2026-0142",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-001"
    },
    {
      "parameter": "Platform",
      "value": "NovaSeq_6000",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-002"
    },
    {
      "parameter": "InstrumentSerial",
      "value": "A01234",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-003"
    },
    {
      "parameter": "FlowCell_Type",
      "value": "S4",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-004"
    },
    {
      "parameter": "Chemistry_Version",
      "value": "v1.5",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-005"
    },
    {
      "parameter": "ReadConfiguration",
      "value": "2x150",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-006"
    },
    {
      "parameter": "RunStartDate",
      "value": "2026-06-07T08:00:00",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-007"
    },
    {
      "parameter": "RunEndDate",
      "value": "2026-06-08T04:00:00",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-008"
    },
    {
      "parameter": "RunDuration_Hours",
      "value": "20",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-009"
    },
    {
      "parameter": "ClusterDensity_K_mm2",
      "value": "950",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-010"
    },
    {
      "parameter": "MinClusterDensity_K_mm2",
      "value": "600",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-011"
    },
    {
      "parameter": "MaxClusterDensity_K_mm2",
      "value": "1200",
      "status": "PASS|WARNING|FAIL",
      "message": "Deterministic rule-based check result",
      "ruleReference": "FS-RULE-012"
    }
  ],
  "criticalFindings": [],
  "warnings": [],
  "audit": {
    "inputHash": "sha256:<calculated at runtime>",
    "rulesVersion": "<utility executable version>",
    "documentation": "SequencerRunQcChecker.documentation.html"
  }
}

Traceability matrix

URSFSTestEvidence
URS-001FS-001, FS-002OQ-001Verify execution and import of valid input.csv.
URS-002FS-005, FS-006OQ-004Repeat the same dataset and compare output.json.
URS-003FS-003, FS-004, FS-010OQ-002, OQ-003Verify missing columns and invalid types.
URS-004FS-005, FS-006OQ-004, PQ-001Verify critical deviations on real/boundary data.
URS-005FS-007, FS-009OQ-005Verify JSON schema and downstream-system suitability.
URS-006FS-008OQ-006Verify identifiers and audit metadata.
URS-007FS-008, FS-010IQ-001, OQ-007Verify documentation completeness and control evidence.
URS-008FS-005, FS-008PQ-002Verify review workflow and no replacement of QA decision.

IQ/OQ/PQ test scenarios

IDScenarioExpected result
IQ-001Verify executable, input.csv, documentation and checksum availability.Delivery set is complete; version is recorded.
OQ-001Valid sample row from input.csv.PASS or acceptable WARNING according to rules.
OQ-002Remove a mandatory CSV column.Schema error or FAIL with missing-column reference.
OQ-003Place a non-numeric value into a numeric field.Type-conversion error with row/field reference.
OQ-004Set a critical parameter outside the limit.FAIL and critical finding.
OQ-005Verify output.json structure.All mandatory sections are present and JSON is valid.
OQ-006Verify batch/sample traceability.Input and result identifiers match.
PQ-001Verify 3–5 real user batches/samples.Result is confirmed by QC/QA review.
PQ-002Verify deviation workflow and manual QA decision.Utility supports review but does not replace approved decision.

QA/QC and change control

  • Do not rename columns without updating validator, documentation and test set.
  • Retain input.csv, output.json, executable version and checksum.
  • Before production use, perform IQ/OQ/PQ or equivalent CSV/CSA verification.
  • Critical limits shall be verified against the approved specification, registration dossier and local SOPs.
  • The utility provides structured QC decision support; final release decision remains with QA/QP and approved procedures.