RheologyProfileValidator

Rheology Profile Validator

dermal fillers endotoxins hyaluronic acid injectables particles rheology sterility
Open selection
Rheology Profile Validator — Filler Rheology Validation (G', G'', tan δ) (ISO 13485)

ℹ️  The utility checks G', tan δ, yield stress, data quality (R²), suitability for injection zone.
ℹ️  Does not use machine learning — rheology rules only.
⚠️  CRITICAL! G' must match injection zone. Tan δ < 0.5 (elasticity). Data in LVE region!

Usage:
  RheologyProfileValidator.exe                            → demo mode
  RheologyProfileValidator.exe input.csv output.json      → evaluate data

Input format:
BatchNumber,G_Prime_Pa,G_DoublePrime_Pa,Complex_Viscosity_Pa_s,Tan_Delta,Crossover_Frequency_Hz,Yield_Stress_Pa,Frequency_Sweep_R2,Temperature_C,Gap_mm,Strain_Percent,Reference_G_Prime_Min,Reference_G_Prime_Max,Target_Application,SterilityTest,Endotoxin_EU_ml

Example (Target: 1=Deep, 2=Mid, 3=Superficial):
RHEO-HA-2026-001,450.0,150.0,25.0,0.33,15.0,25.0,0.995,25.0,1.0,0.5,400.0,600.0,1,0,0.1

— WHY IS THIS NEEDED?
Rheology determines clinical behavior of filler:
• G' (Elastic Modulus) — ability to restore shape, "lifting capacity". High G' for deep zones, low for superficial.
• G'' (Loss Modulus) — viscous component, resistance to flow.
• Tan δ (G''/G') — ratio of viscosity to elasticity. For fillers preferably < 0.5 (elasticity dominance).
• Yield Stress — force required to start flow. Important for needle extrusion and tissue stability.
• Crossover Frequency — point where gel transitions from elastic to viscous behavior.
⚠️ CRITICAL:
• G' Match to Injection Zone — error leads to product migration or lumps.
• Tan δ < 0.5 — ensures longevity and resistance to deformation by facial muscles.
• Data Quality (R² ≥ 0.98) — guarantee that measurements performed in Linear Viscoelastic Region (LVE).
• Temperature 25°C (or 37°C) — standardization mandatory as viscosity highly depends on T.
• Sterility and Endotoxins — mandatory for any injectable forms.
⚠️ Note: Utility not just compares numbers but assesses product suitability for intended purpose (Deep/Mid/Superficial). E.g., filler with G'=100 Pa unsuitable for cheek augmentation (need >400 Pa) but ideal for lips or fine lines. Checking frequency sweep R² guarantees operator did not exceed linear deformation region where moduli are constant.

input.csv

BatchNumber,G_Prime_Pa,G_DoublePrime_Pa,Complex_Viscosity_Pa_s,Tan_Delta,Crossover_Frequency_Hz,Yield_Stress_Pa,Frequency_Sweep_R2,Temperature_C,Gap_mm,Strain_Percent,Reference_G_Prime_Min,Reference_G_Prime_Max,Target_Application,SterilityTest,Endotoxin_EU_ml
RHEO-HA-2026-001,450.0,150.0,25.0,0.33,15.0,25.0,0.995,25.0,1.0,0.5,400.0,600.0,1,0,0.1

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: RheologyProfileValidatorAPI / object: Rheology Profile ValidatorCategory: Dermal fillers

1. Scope

The Rheology Profile Validator utility is used for Rheology Profile Validator. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Dermal fillers

Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility

2. Execution modes

RheologyProfileValidator.exe                            → demo mode
RheologyProfileValidator.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • microbiology and pathogens
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ CRITICAL! G' must match injection zone. Tan δ < 0.5 (elasticity). Data in LVE region!
  • Tan δ (G''/G') — ratio of viscosity to elasticity. For fillers preferably < 0.5 (elasticity dominance).
  • ⚠️ CRITICAL:
  • G' Match to Injection Zone — error leads to product migration or lumps.
  • Tan δ < 0.5 — ensures longevity and resistance to deformation by facial muscles.
  • Data Quality (R² ≥ 0.98) — guarantee that measurements performed in Linear Viscoelastic Region (LVE).
  • Temperature 25°C (or 37°C) — standardization mandatory as viscosity highly depends on T.
  • Sterility and Endotoxins — mandatory for any injectable forms.
  • ⚠️ Note: Utility not just compares numbers but assesses product suitability for intended purpose (Deep/Mid/Superficial). E.g., filler with G'=100 Pa unsuitable for cheek augmentation (need >400 Pa) but ideal for lips or fine lines. Checking frequency sweep R² guarantees operator did not exceed linear deformation region where moduli are constant.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Rheology Profile Validator with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: microbiology and pathogens, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberRHEO-HA-2026-001mandatory field; used for batch/lot traceability
2G_Prime_Padecimalas specifiedG Prime Pa450.0mandatory numeric value; rule comparison is performed by the utility
3G_DoublePrime_Padecimalas specifiedG Double Prime Pa150.0mandatory numeric value; rule comparison is performed by the utility
4Complex_Viscosity_Pa_sdecimalas specifiedComplex Viscosity Pa s25.0mandatory numeric value; rule comparison is performed by the utility
5Tan_Deltadecimalas specifiedTan Delta0.33mandatory numeric value; rule comparison is performed by the utility
6Crossover_Frequency_Hzdecimalas specifiedCrossover Frequency Hz15.0mandatory numeric value; rule comparison is performed by the utility
7Yield_Stress_Padecimalas specifiedYield Stress Pa25.0mandatory numeric value; rule comparison is performed by the utility
8Frequency_Sweep_R2decimalas specifiedFrequency Sweep R20.995mandatory numeric value; rule comparison is performed by the utility
9Temperature_Cdecimal°CTemperature C25.0mandatory numeric value; rule comparison is performed by the utility
10Gap_mmdecimalas specifiedGap mm1.0mandatory numeric value; rule comparison is performed by the utility
11Strain_Percentdecimal%Strain %0.5mandatory numeric value; rule comparison is performed by the utility
12Reference_G_Prime_MindecimalminReference G Prime Min400.0mandatory numeric value; rule comparison is performed by the utility
13Reference_G_Prime_Maxdecimalas specifiedReference G Prime Max600.0mandatory numeric value; rule comparison is performed by the utility
14Target_Applicationdecimalas specifiedTarget Application1mandatory numeric value; rule comparison is performed by the utility
15SterilityTestboolean / flag0/1Sterility Test0valid flag required; prohibited organisms and growth are normally expected as 0 / absent
16Endotoxin_EU_mldecimalas specifiedEndotoxin EU ml0.1numeric value; critical safety attribute compared with endotoxin limit

CSV example

BatchNumber,G_Prime_Pa,G_DoublePrime_Pa,Complex_Viscosity_Pa_s,Tan_Delta,Crossover_Frequency_Hz,Yield_Stress_Pa,Frequency_Sweep_R2,Temperature_C,Gap_mm,Strain_Percent,Reference_G_Prime_Min,Reference_G_Prime_Max,Target_Application,SterilityTest,Endotoxin_EU_ml
RHEO-HA-2026-001,450.0,150.0,25.0,0.33,15.0,25.0,0.995,25.0,1.0,0.5,400.0,600.0,1,0,0.1

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable RheologyProfileValidator.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Rheology Profile Validator, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "RheologyProfileValidator",
  "api": "Rheology Profile Validator",
  "batchNumber": "RHEO-HA-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "RHEO-HA-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "G_Prime_Pa",
      "value": "450.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "G_DoublePrime_Pa",
      "value": "150.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Complex_Viscosity_Pa_s",
      "value": "25.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Tan_Delta",
      "value": "0.33",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Crossover_Frequency_Hz",
      "value": "15.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Yield_Stress_Pa",
      "value": "25.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Frequency_Sweep_R2",
      "value": "0.995",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-005Positive pathogen/microbiological flag or microbiological limit excursionCritical FAIL is produced with the offending parameter.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.