PolymethylMethacrylateQualityChecker

Polymethyl Methacrylate

dermal fillers endotoxins hyaluronic acid particles rheology sterility
Open selection
PMMA Filler Quality Checker — Permanent Polymethylmethacrylate Fillers (Bellafill type) (ISO 13485, EU MDR)

ℹ️  The utility checks microsphere size (30-50 µm), no sharp edges, residual MMA monomer, PMMA content, sterility.
ℹ️  Does not use machine learning — rule-based logic only.
⚠️  CRITICAL! Sharp edges = 0. MMA Monomer ≤ 10 ppm. Particle size strictly 30-50 µm!

Usage:
  PolymethylMethacrylateQualityChecker.exe                            → demo mode
  PolymethylMethacrylateQualityChecker.exe input.csv output.json      → evaluate data

Input format:
BatchNumber,PMMA_Content_Percent,Microsphere_Size_D50_um,Microsphere_Size_Min_um,Microsphere_Size_Max_um,Sharp_Edges_Count,Agglomerates_Percent,Residual_MMA_Monomer_ppm,Carrier_Gel_Type,Carrier_Gel_Concentration_mgPerMl,Viscosity_Pa_s,pH_Value,SterilityTest,Endotoxin_EU_ml,Protein_Residue_ppm

Example:
PMMA-BELLA-2026-001,20.0,40.0,32.0,55.0,0,0.1,5.0,1,35.0,25.0,7.1,0,0.1,8.0

— WHY IS THIS NEEDED?
Permanent Polymethylmethacrylate (PMMA) fillers — long-acting products:
• Bellafill (Artefill), Silikon-1000 (off-label), etc.
• PMMA microspheres (30-50 µm) suspended in carrier (bovine collagen or HA)
• Microspheres do not degrade, creating permanent scaffold for own collagen
• Require strict control of shape and size for safety
⚠️ CRITICAL:
• Microsphere Size 30–50 µm — critical range: <30 µm phagocytosed by macrophages (disappear), >60 µm may migrate or be visible under skin
• Sharp Edges = 0 — any surface irregularities cause tissue trauma, inflammation, and granulomas. Only perfect spheres allowed!
• Residual MMA Monomer ≤ 10 ppm — methyl methacrylate is toxic, causes necrosis and severe inflammation
• Agglomerates ≤ 0.5% — clumped microspheres hard to inject and create lumps
• Protein (for collagen carrier) ≤ 10 ppm — anaphylaxis risk (skin test required before use)
• Sterility and Endotoxins — MANDATORY for implantable products
⚠️ Note: PMMA microspheres synthesized by emulsion polymerization. Critically important to control sphericity and surface smoothness (microscopy). Carrier (collagen) resorbs over time, leaving only PMMA microspheres. Due to permanent effect, safety requirements higher than for biodegradable fillers. Any deviation in particle shape unacceptable.

input.csv

BatchNumber,PMMA_Content_Percent,Microsphere_Size_D50_um,Microsphere_Size_Min_um,Microsphere_Size_Max_um,Sharp_Edges_Count,Agglomerates_Percent,Residual_MMA_Monomer_ppm,Carrier_Gel_Type,Carrier_Gel_Concentration_mgPerMl,Viscosity_Pa_s,pH_Value,SterilityTest,Endotoxin_EU_ml,Protein_Residue_ppm
PMMA-BELLA-2026-001,20.0,40.0,32.0,55.0,0,0.1,5.0,1,35.0,25.0,7.1,0,0.1,8.0

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: PolymethylMethacrylateQualityCheckerAPI / object: Polymethyl MethacrylateCategory: Dermal fillers

1. Scope

The Polymethyl Methacrylate Quality Checker utility is used for Polymethyl Methacrylate. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Dermal fillers

Tags: dermal fillers, endotoxins, hyaluronic acid, particles, rheology, sterility

2. Execution modes

PolymethylMethacrylateQualityChecker.exe                            → demo mode
PolymethylMethacrylateQualityChecker.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • assay / content
  • microbiology and pathogens
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ CRITICAL! Sharp edges = 0. MMA Monomer ≤ 10 ppm. Particle size strictly 30-50 µm!
  • ⚠️ CRITICAL:
  • Microsphere Size 30–50 µm — critical range: <30 µm phagocytosed by macrophages (disappear), >60 µm may migrate or be visible under skin
  • Sharp Edges = 0 — any surface irregularities cause tissue trauma, inflammation, and granulomas. Only perfect spheres allowed!
  • Residual MMA Monomer ≤ 10 ppm — methyl methacrylate is toxic, causes necrosis and severe inflammation
  • Agglomerates ≤ 0.5% — clumped microspheres hard to inject and create lumps
  • Protein (for collagen carrier) ≤ 10 ppm — anaphylaxis risk (skin test required before use)
  • Sterility and Endotoxins — MANDATORY for implantable products
  • ⚠️ Note: PMMA microspheres synthesized by emulsion polymerization. Critically important to control sphericity and surface smoothness (microscopy). Carrier (collagen) resorbs over time, leaving only PMMA microspheres. Due to permanent effect, safety requirements higher than for biodegradable fillers. Any deviation in particle shape unacceptable.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Polymethyl Methacrylate with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content, microbiology and pathogens, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberPMMA-BELLA-2026-001mandatory field; used for batch/lot traceability
2PMMA_Content_Percentdecimal%PMMA Content %20.0numeric value; compared with assay/purity limit from specification or method
3Microsphere_Size_D50_umdecimalµmMicrosphere Size D50 um40.0mandatory numeric value; rule comparison is performed by the utility
4Microsphere_Size_Min_umdecimalµmMicrosphere Size Min um32.0mandatory numeric value; rule comparison is performed by the utility
5Microsphere_Size_Max_umdecimalµmMicrosphere Size Max um55.0mandatory numeric value; rule comparison is performed by the utility
6Sharp_Edges_Countdecimalas specifiedSharp Edges Count0numeric value; compared with microbiological limits
7Agglomerates_Percentdecimal%Agglomerates %0.1mandatory numeric value; rule comparison is performed by the utility
8Residual_MMA_Monomer_ppmdecimalppmResidual MMA Monomer ppm5.0mandatory numeric value; rule comparison is performed by the utility
9Carrier_Gel_Typedecimalas specifiedCarrier Gel Type1mandatory numeric value; rule comparison is performed by the utility
10Carrier_Gel_Concentration_mgPerMldecimalmg/mlCarrier Gel Concentration mg per Ml35.0numeric value; compared with assay/purity limit from specification or method
11Viscosity_Pa_sdecimalas specifiedViscosity Pa s25.0mandatory numeric value; rule comparison is performed by the utility
12pH_ValuedecimalpHp H Value7.1mandatory numeric value; rule comparison is performed by the utility
13SterilityTestboolean / flag0/1Sterility Test0valid flag required; prohibited organisms and growth are normally expected as 0 / absent
14Endotoxin_EU_mldecimalas specifiedEndotoxin EU ml0.1numeric value; critical safety attribute compared with endotoxin limit
15Protein_Residue_ppmdecimalppmProtein Residue ppm8.0mandatory numeric value; rule comparison is performed by the utility

CSV example

BatchNumber,PMMA_Content_Percent,Microsphere_Size_D50_um,Microsphere_Size_Min_um,Microsphere_Size_Max_um,Sharp_Edges_Count,Agglomerates_Percent,Residual_MMA_Monomer_ppm,Carrier_Gel_Type,Carrier_Gel_Concentration_mgPerMl,Viscosity_Pa_s,pH_Value,SterilityTest,Endotoxin_EU_ml,Protein_Residue_ppm
PMMA-BELLA-2026-001,20.0,40.0,32.0,55.0,0,0.1,5.0,1,35.0,25.0,7.1,0,0.1,8.0

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable PolymethylMethacrylateQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Polymethyl Methacrylate, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "PolymethylMethacrylateQualityChecker",
  "api": "Polymethyl Methacrylate",
  "batchNumber": "PMMA-BELLA-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "PMMA-BELLA-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "PMMA_Content_Percent",
      "value": "20.0",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Microsphere_Size_D50_um",
      "value": "40.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Microsphere_Size_Min_um",
      "value": "32.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Microsphere_Size_Max_um",
      "value": "55.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Sharp_Edges_Count",
      "value": "0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Agglomerates_Percent",
      "value": "0.1",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Residual_MMA_Monomer_ppm",
      "value": "5.0",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-005Positive pathogen/microbiological flag or microbiological limit excursionCritical FAIL is produced with the offending parameter.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.