PET/CT and Dose-Calibrator Cross-Calibration

Compares dose-calibrator activity with activity concentration measured from the PET phantom image.

  • decay correction between calibration and scan;
  • expected activity concentration;
  • recovery/cross-calibration ratio and deviation from unity.

The formula and tolerance shall match the approved phantom protocol, radionuclide and local quantitative-QC programme.

input.csv

StudyID,PhantomType,DoseCalibratorActivityMBq,CalibrationTime,ScanStartTime,MeasuredActivityConcentrationBQ_per_ML,PhantomVolumeML,PatientWeightKg,DecayCorrectedDoseMBq
QC-PET-2026-001,NEMA_Image_Quality,100.0,2026-07-22 08:00:00,2026-07-22 09:00:00,28500.0,3000.0,0.0,0.0
QC-PET-2026-002,Jaszczak_Photon,50.0,2026-07-22 10:00:00,2026-07-22 10:30:00,14000.0,1500.0,0.0,0.0
QC-PET-2026-003,Cylinder_Uniform,80.0,2026-07-22 12:00:00,2026-07-22 13:00:00,20000.0,4000.0,0.0,0.0

URS & FS — user requirements and functional specification

URS — User Requirements Specification

  • The utility shall accept input.csv with the headers defined by the data contract.
  • Before calculation, the utility shall detect missing columns, type errors, empty mandatory values and illogical dates or ranges.
  • The utility shall deterministically perform “PET/CT and Dose-Calibrator Cross-Calibration”.
  • Limits and rules shall be loaded from an approved profile or explicitly fixed in the validated version.
  • Each record shall receive a clear final status: PASS, WARNING or FAIL, with the reason.
  • The result shall retain source values, derived values, applied limits and individual check statuses.
  • Invalid input shall not return a successful result. The output supports qualified review under an approved procedure.

FS — Functional Specification

  1. Run PETCTCrossCalibrationChecker.exe in demonstration mode or with input.csv output.json arguments.
  2. Read UTF-8 CSV and map columns to the data contract.
  3. Convert values to the expected types and validate mandatory fields.
  4. Perform calculations and rule checks for “PET/CT and Dose-Calibrator Cross-Calibration”.
  5. Create individual checks and derive the overall status from the worst result.
  6. Write machine-readable output.json; populate errors when a valid calculation cannot be produced.

input.csv example

StudyID,PhantomType,DoseCalibratorActivityMBq,CalibrationTime,ScanStartTime,MeasuredActivityConcentrationBQ_per_ML,PhantomVolumeML,PatientWeightKg,DecayCorrectedDoseMBq
QC-PET-2026-001,NEMA_Image_Quality,100.0,2026-07-22 08:00:00,2026-07-22 09:00:00,28500.0,3000.0,0.0,0.0
QC-PET-2026-002,Jaszczak_Photon,50.0,2026-07-22 10:00:00,2026-07-22 10:30:00,14000.0,1500.0,0.0,0.0
QC-PET-2026-003,Cylinder_Uniform,80.0,2026-07-22 12:00:00,2026-07-22 13:00:00,20000.0,4000.0,0.0,0.0

Minimum output.json structure

{
  "utility": "PETCTCrossCalibrationChecker",
  "source": "input.csv",
  "status": "PASS|WARNING|FAIL",
  "derivedValues": {},
  "checks": [
    {
      "parameter": "example",
      "value": 0,
      "limit": "configured profile",
      "status": "PASS",
      "message": "criterion satisfied"
    }
  ],
  "errors": []
}

Minimum OQ scenarios

  • A row within the approved profile shall return PASS.
  • A non-critical limit violation shall return WARNING or a review status.
  • A critical violation shall return FAIL or the applicable operational decision.
  • A missing column, invalid type or impossible date shall populate errors and shall not return PASS.

Before operational use

  • Reconcile formulas and limits with manufacturer documentation and the approved SOP.
  • Fix the scanner model, software version and profile version.
  • After an executable, formula or profile change, repeat affected tests.
  • Retain the source CSV and JSON result together.

Included in packages

PET Scanner State QC Suite

14 utilities for PET/PET-CT scanner state control: Ge-68/Ga-68 source activity, blank sinogram, detector efficiency, normalization acquisition and factor distribution, normalization-file assignment, trending, energy window, TOF, attenuation-correction map, PET/CT cross-calibration and the daily operational decision.

Open

RPH use case: PET scanner state, gamma camera and radiation physics

The linked medical-device/radiation-physics layer: PET detector QC, normalization, AC map, cross-calibration, TOF, gamma-camera uniformity, source strength and staff dose.

Open

RPH workflow: instruments, devices and radiation safety

PET/gamma-camera state, calibration, detector QC, energy window, TOF, shielding and radiation safety.

Open

RPH workflow: stability, storage, decay and waste

Kinetic stability, shelf life, decay correction, expiry, storage, trend and radioactive waste.

Open