NMR Structure Quality Checker — Control of NMR Spectrum Quality

ℹ️  Utility checks critical parameters of Nuclear Magnetic Resonance spectra quality (USP <761> / Ph. Eur. 2.2.33):
   • Signal-to-Noise Ratio (S/N): ≥150:1 (for quantitative analysis)
   • Resolution: ≤0.55 Hz (at half height)
   • Linewidth: ≤1.0 Hz (at 0.55% height, shimming indicator)
   • Shift Calibration: standard deviation ≤0.01 ppm

⚠️  CRITICAL: Poor shimming (broad peaks) makes multiplet separation and accurate integration impossible!
   Low S/N hides minor impurities or spin-spin coupling.

Usage:
 NMRStructureQualityChecker.exe                            → demo mode (console output)
 NMRStructureQualityChecker.exe input.csv output.json      → evaluate your data

Input format:
BatchNumber,SignalToNoiseRatio,ResolutionHz,Linewidth055Hz,ReferenceShiftDeviation,PhaseCorrectionScore

Example:
 NMR-STR-2026-001,250.0,0.45,0.8,0.002,98.0

— WHY IS THIS NEEDED?
NMR is the most informative method for confirming structure of organic molecules and APIs.
• Spectrum quality directly affects reliability of structure elucidation and quantitative calculations (qNMR).
• Resolution and linewidth parameters reflect magnetic field homogeneity (shimming quality).
• S/N determines impurity detection limit and signal integration reliability.
• Automation of QC parameter check allows operator to immediately re-acquire spectrum if poor, saving time.

⚠️  CRITICAL:
• Shimming must be optimized before each measurement.
• Sample temperature must stabilize before acquisition.
• Sample concentration must be sufficient for good S/N without detector saturation.
• Solvent must be high-quality deuterated.

Key features:
• Assessment of spectrum suitability for structural identification.
• Control of instrument technical parameters (magnet, probe).
• Support for Bruker, Jeol, Agilent spectrometer data.

Critical parameters:
• S/N Ratio: >= 150
• Resolution: <= 0.55 Hz
• Linewidth (0.55%): <= 1.0 Hz
• Reference Shift: <= 0.01 ppm

💡 Usage tips:
1. Use standard sample (e.g., ethylbenzene in CDCl3) for daily resolution and sensitivity check.
2. Monitor cryogen levels (liquid nitrogen/helium) in magnet.
3. Avoid paramagnetic impurities (oxygen, metals) which broaden peaks.
4. For quantitative NMR (qNMR), use sufficient relaxation delay (D1 >= 5*T1).
5. Check phase and baseline correction before integration.

⚠️ Note: NMR requires high stability of external conditions. Vibrations and temperature fluctuations in the lab can significantly degrade spectrum resolution.

input.csv

BatchNumber,SN_Ratio,Resolution_Hz,Linewidth_055_Hz,RefShift_Dev_ppm,Phase_Score
NMR-STR-2026-001,250.0,0.45,0.8,0.002,98.0
NMR-STR-2026-002,180.0,0.50,0.9,0.005,95.0
NMR-FAIL-SHIM-003,120.0,1.20,2.5,0.001,90.0

URS & FS — NMR Structure

This document describes the user requirements and functional specification for NMRStructureQualityChecker. The utility is intended for QC laboratory use as a rule-based check of data prepared from an instrument, LIMS, ELN, MES or an approved input.csv.

URS — User Requirements Specification

IDRequirementCriticalityAcceptance criterion
URS-001The utility shall accept input.csv with approved columns: BatchNumber, SN_Ratio, Resolution_Hz, Linewidth_055_Hz, RefShift_Dev_ppm, Phase_Score.HighThe CSV file is processed without manual header editing.
URS-002The utility shall perform deterministic evaluation for “NMR Structure” without machine learning.HighThe same input data produce the same JSON result.
URS-003The utility shall validate mandatory fields, numeric formats, flags, ranges and domain plausibility.HighSchema and conversion errors are explicitly reported.
URS-004The utility shall generate output.json with PASS / WARNING / FAIL statuses, source values, warnings and failures.HighThe JSON result is suitable for review, deviation investigation and integration.
URS-005The documentation shall support IQ/OQ/PQ or CSV/CSA verification.MediumURS/FS, CSV/JSON contract and test scenarios are supplied with the utility.

input.csv contract

#FieldSamplePurpose
1BatchNumberNMR-STR-2026-001Input parameter for deterministic QC evaluation.
2SN_Ratio250.0Input parameter for deterministic QC evaluation.
3Resolution_Hz0.45Input parameter for deterministic QC evaluation.
4Linewidth_055_Hz0.8Input parameter for deterministic QC evaluation.
5RefShift_Dev_ppm0.002Input parameter for deterministic QC evaluation.
6Phase_Score98.0Input parameter for deterministic QC evaluation.

FS — Functional Specification

IDFunctionImplementation
FS-001CSV importRead input.csv; validate header, column count and encoding.
FS-002Schema validationCheck required fields and permitted input values.
FS-003Rule engineApply domain rules, limits and system suitability/specification checks described in the source utility description.
FS-004Status aggregationProduce final status: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance.
FS-005JSON exportWrite machine-readable output.json for LIMS/ELN/MES and QA/QC review.

OQ/PQ scenarios

  • OQ-001: a valid sample row shall be processed without schema error.
  • OQ-002: a missing mandatory column shall produce a schema error.
  • OQ-003: a non-numeric value in a numeric field shall produce a conversion error.
  • OQ-004: a critical parameter outside the limit shall produce FAIL or a critical finding.
  • PQ-001: user real batches shall be checked with retention of input.csv, output.json, utility version and checksum.

Included in packages

Immunology QC Suite

QC utility package for immunology: monoclonal antibodies, biosimilars, immunomodulators, immunosuppressants, ELISA/SPR, HCP, protein characterization, sterility and pyrogen/release checks.

Open

LabEx QC Suite

LabEx package for laboratory QC methods: chromatography, spectroscopy, microbiology, sterility, endotoxins, particles, dissolution, water, cleanrooms and sample preparation.

Open

Radiology QC Suite

QC package for radiology and diagnostic imaging: radiopharmaceuticals, PET/SPECT, contrast media, iodinated and gadolinium products, plus particle, sterility and endotoxin checks.

Open

Toxicology QC Suite

QC utility package for toxicologically relevant risks: genotoxic impurities, residual solvents, elemental impurities, heavy metals, mercury, mycotoxins, pesticides, extractables/leachables and degradation impurities.

Open