NiacinamideQualityChecker

Niacinamide

assay heavy metals impurities minerals premix specification vitamins water quality
Open selection

Niacinamide Quality Checker — Vitamin B3, Nicotinamide (Ph. Eur. 2.8.1)

ℹ️ The utility checks assay, specific rotation, melting point, pH, impurities (nicotinic acid), inorganic impurities.
ℹ️ Does not use machine learning — rule-based logic only.
⚠️ Critical for vitamin products, dietary supplements, cosmetics, parenteral nutrition.

Usage:
NiacinamideQualityChecker.exe → demo mode (console output)
NiacinamideQualityChecker.exe input.csv output.json → evaluate your data

Input format:
BatchNumber,AssayPercent,SpecificRotation,MeltingPoint,pH,LossOnDrying,HeavyMetals,NicotinicAcid,OtherImpurities,TotalImpurities,Chlorides,Sulfates,Fluorides,ClarityNTU,ColorAPHA,EthanolPPM,AcetonePPM

Example:
NIA-2026-001,99.5,0.0,129.5,6.0,0.5,10,0.2,0.3,0.5,0.03,0.02,30,2,30,800,500

— WHY IS THIS NEEDED?
Niacinamide (vitamin B3, nicotinamide) is a water-soluble vitamin critical for:
• Carbohydrate, fat, and protein metabolism (NAD/NADP coenzyme)
• Skin health (anti-inflammatory action, barrier function)
• Nervous system and DNA synthesis
• Vitamin complexes, dietary supplements, parenteral nutrition
• Cosmetic products (pigmentation reduction, hydration)
⚠️ Must distinguish from nicotinic acid (niacin) — different pharmacological properties (niacin causes "flushing")!
Critical parameters:
• Assay 98.0–102.0% (on dried basis)
• Specific rotation -0.5° to +0.5° (optically inactive, but checked for identification)
• Melting point 128–131°C (key identification parameter)
• pH of 2% solution: 5.0–7.0
• Nicotinic acid ≤0.5% (critical impurity — different drug!)
• Total impurities ≤1.0%
• Loss on drying ≤1.0%
• Heavy metals ≤20 ppm
• Chlorides ≤0.05%, sulfates ≤0.05%, fluorides ≤50 ppm
• Clarity ≤5 NTU, color ≤50 APHA
Monograph compliance is mandatory for registration. Stable but hygroscopic — requires moisture protection!

input.csv

BatchNumber,AssayPercent,SpecificRotation,MeltingPoint,pH,LossOnDrying,HeavyMetals,NicotinicAcid,OtherImpurities,TotalImpurities,Chlorides,Sulfates,Fluorides,ClarityNTU,ColorAPHA,EthanolPPM,AcetonePPM
NIA-2026-001,99.5,0.0,129.5,6.0,0.5,10,0.2,0.3,0.5,0.03,0.02,30,2,30,800,500

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: NiacinamideQualityCheckerAPI / object: NiacinamideCategory: Vitamins and minerals

1. Scope

The Niacinamide Quality Checker utility is used for Niacinamide. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Vitamins and minerals

Tags: assay, heavy metals, impurities, minerals, premix, specification, vitamins, water quality

2. Execution modes

NiacinamideQualityChecker.exe                            → demo mode (console output)
NiacinamideQualityChecker.exe input.csv output.json      → evaluate your data

3. Key controlled areas

  • assay / content
  • residual solvents
  • heavy metals / elemental impurities
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ Critical for vitamin products, dietary supplements, cosmetics, parenteral nutrition.
  • ⚠️ Must distinguish from nicotinic acid (niacin) — different pharmacological properties (niacin causes "flushing")!
  • Critical parameters:
  • Assay 98.0–102.0% (on dried basis)
  • Specific rotation -0.5° to +0.5° (optically inactive, but checked for identification)
  • Melting point 128–131°C (key identification parameter)
  • pH of 2% solution: 5.0–7.0
  • Nicotinic acid ≤0.5% (critical impurity — different drug!)
  • Total impurities ≤1.0%
  • Loss on drying ≤1.0%
  • Heavy metals ≤20 ppm
  • Chlorides ≤0.05%, sulfates ≤0.05%, fluorides ≤50 ppm
  • Clarity ≤5 NTU, color ≤50 APHA
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Niacinamide with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content, residual solvents, heavy metals / elemental impurities, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberNIA-2026-001mandatory field; used for batch/lot traceability
2AssayPercentdecimal%Assay %99.5numeric value; compared with assay/purity limit from specification or method
3SpecificRotationdecimalas specifiedSpecific Rotation0.0mandatory numeric value; rule comparison is performed by the utility
4MeltingPointdecimal°CMelting Point129.5mandatory numeric value; rule comparison is performed by the utility
5pHdecimalpHp H6.0mandatory numeric value; rule comparison is performed by the utility
6LossOnDryingdecimalas specifiedLoss On Drying0.5mandatory numeric value; rule comparison is performed by the utility
7HeavyMetalsdecimalas specifiedHeavy Metals10numeric value; checked as elemental/metal impurity
8NicotinicAciddecimalas specifiedNicotinic Acid0.2mandatory numeric value; rule comparison is performed by the utility
9OtherImpuritiesdecimalas specifiedOther Impurities0.3mandatory numeric value; rule comparison is performed by the utility
10TotalImpuritiesdecimalas specifiedTotal Impurities0.5mandatory numeric value; rule comparison is performed by the utility
11Chloridesdecimalas specifiedChlorides0.03mandatory numeric value; rule comparison is performed by the utility
12Sulfatesdecimalas specifiedSulfates0.02mandatory numeric value; rule comparison is performed by the utility
13Fluoridesdecimalas specifiedFluorides30mandatory numeric value; rule comparison is performed by the utility
14ClarityNTUdecimalas specifiedClarity NTU2mandatory numeric value; rule comparison is performed by the utility
15ColorAPHAdecimalas specifiedColor APHA30mandatory numeric value; rule comparison is performed by the utility
16EthanolPPMdecimalppmEthanol PPM800numeric value; checked as residual solvent against the applicable limit
17AcetonePPMdecimalppmAcetone PPM500numeric value; checked as residual solvent against the applicable limit

CSV example

BatchNumber,AssayPercent,SpecificRotation,MeltingPoint,pH,LossOnDrying,HeavyMetals,NicotinicAcid,OtherImpurities,TotalImpurities,Chlorides,Sulfates,Fluorides,ClarityNTU,ColorAPHA,EthanolPPM,AcetonePPM
NIA-2026-001,99.5,0.0,129.5,6.0,0.5,10,0.2,0.3,0.5,0.03,0.02,30,2,30,800,500

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable NiacinamideQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Niacinamide, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "NiacinamideQualityChecker",
  "api": "Niacinamide",
  "batchNumber": "NIA-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "NIA-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "AssayPercent",
      "value": "99.5",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "SpecificRotation",
      "value": "0.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "MeltingPoint",
      "value": "129.5",
      "unit": "°C",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "pH",
      "value": "6.0",
      "unit": "pH",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "LossOnDrying",
      "value": "0.5",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "HeavyMetals",
      "value": "10",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "NicotinicAcid",
      "value": "0.2",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-007Residual solvent above applicable limitFAIL is produced with solvent name and measured value.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Vitamins & Minerals QC Suite

Quality-control package for vitamins, minerals, premixes and nutrient substances.

Open