InfraLUM_ID_Ribavirin

InfraLUM ID Ribavirin

Lumex QC URS & FS input.csv output.json rule-based LIMS-ready InfraLUM
Open selection

Utility description: InfraLUM ID Ribavirin

InfraLUM ID Ribavirin — Ribavirin Identification by IR Spectroscopy

ℹ️  Utility checks IR spectrum compliance with reference (Ph. Eur. / Russian Pharmacopoeia):
    • Correlation coefficient with reference spectrum: ≥0.995
    • Peak 3400 cm⁻¹ (O-H/N-H stretching, hydrogen bonds): ±25 cm⁻¹
    • Peak 1680 cm⁻¹ (carboxamide C=O): ±10 cm⁻¹ — AMIDE MARKER
    • Peak 1620 cm⁻¹ (triazole C=N/NH₂): ±10 cm⁻¹
    • Peak 1210 cm⁻¹ (ribose C-O-C): ±10 cm⁻¹ — SUGAR MARKER
    • Peak 1050 cm⁻¹ (primary alcohol C-O CH₂OH): ±10 cm⁻¹
    • Baseline noise: ≤0.005 Abs

⚠️  IMPORTANT: Ribavirin is a synthetic guanosine nucleoside analogue.
    Instead of a purine base, the molecule contains 1,2,4-triazole-3-carboxamide
    linked to β-D-ribofuranose via N-glycosidic bond.
    The triazole ring mimics the imidazole portion of purine, while the carboxamide
    group (-CONH₂) replaces guanine's amino group.
    Multiple ribose hydroxyl groups form an extensive hydrogen bonding network,
    creating a very broad absorption band at 3200–3500 cm⁻¹.

Usage:
  InfraLUM_ID_Ribavirin.exe                            → demo mode (console output)
  InfraLUM_ID_Ribavirin.exe input.csv output.json      → evaluate your data

Input format:
BatchNumber,CorrelationCoefficient,Peak3400_cm1,Peak1680_cm1,Peak1620_cm1,Peak1210_cm1,Peak1050_cm1,BaselineNoiseAbs,PreparationMethod

Example (PreparationMethod: KBr / ATR / Nujol):
  RBV-ID-2026-027,0.9994,3395.0,1678.0,1622.0,1208.0,1052.0,0.0014,KBr

— WHY IS THIS NEEDED?
Ribavirin is a broad-spectrum antiviral agent.
• Used in combination with interferon or sofosbuvir for hepatitis C treatment.
• Mechanism: viral RNA polymerase inhibition and intracellular GTP pool depletion.
• IR spectroscopy is the primary identity confirmation method per Ph. Eur. and Russian Pharmacopoeia.
• Reference spectrum established in Ph. Eur. Ribavirin and Russian Pharmacopoeia FS.2.4.00270.18.
• InfraLUM FT-08/FT-10 instruments are widely used in Russian QC laboratories.

⚠️  CRITICAL:
• Correlation ≥0.995 — strict matching criterion with reference.
• Peak 1680 cm⁻¹ (amide C=O) — primary carboxamide group marker.
  Distinguishes ribavirin from natural nucleosides (guanosine has no amide).
  Shift to ~1650 cm⁻¹ may indicate amide hydrolysis to acid.
• Peak 1620 cm⁻¹ (triazole C=N + NH₂ bend) — heterocyclic base marker.
  Absence indicates triazole ring destruction.
• Peak 1210 cm⁻¹ (ribose C-O-C) — specific furanose ring marker.
  Distinguishes ribavirin from deoxyribonucleosides (different sugar conformation).
• Peak 1050 cm⁻¹ (C-O CH₂OH) — confirms primary alcohol group presence.
• Broad band 3200–3500 cm⁻¹ — crystal structure "fingerprint".
• Noise >0.005 Abs renders spectrum unsuitable for identification.

Key features:
• Uses NAMED TUPLE (record struct EvalResult) —
  eliminates positional field access errors.
• Specification based on Ph. Eur. Ribavirin and Russian Pharmacopoeia FS.2.4.00270.18.
• Control of five characteristic peaks + correlation + spectrum quality.
• Differentiation from natural nucleosides via amide C=O peak.
• Full LabWare LIMS compatibility.

Critical parameters:
• Correlation Coefficient: ≥0.995
• Peak 3400 cm⁻¹ (O-H/N-H H-bonded): 3400 ± 25 cm⁻¹
• Peak 1680 cm⁻¹ (carboxamide C=O): 1680 ± 10 cm⁻¹
• Peak 1620 cm⁻¹ (triazole C=N/NH₂): 1620 ± 10 cm⁻¹
• Peak 1210 cm⁻¹ (ribose C-O-C): 1210 ± 10 cm⁻¹
• Peak 1050 cm⁻¹ (primary alcohol C-O): 1050 ± 10 cm⁻¹
• Baseline Noise: ≤0.005 Abs

💡 Usage tips:
1. KBr tablets must be DRY — pre-dry at 120°C before use.
   Ribavirin is VERY hygroscopic! Moisture radically distorts the 3200–3500 cm⁻¹ region.
2. Sample concentration in KBr: 0.5–1.0% (w/w).
3. Record spectrum in 4000–400 cm⁻¹ range, 4 cm⁻¹ resolution, 32 scans.
4. Record background spectrum of pure KBr before each measurement.
5. For ATR method, ensure tight contact; ribavirin compresses well.
6. Reference spectrum should be obtained on the same instrument under identical conditions.
7. Prepare sample QUICKLY under low humidity conditions (<40% RH).

⚠️ Note: Ribavirin exists in two anomeric forms (α and β), but only the β-D-ribofuranose
    form is pharmacopoeial. The α-anomer shows differences in the ribose C-O-C region
    (1000–1250 cm⁻¹) and may reduce correlation.
    If α-anomer presence is suspected, perform specific rotation test or chiral HPLC.
    The carboxamide group (-CONH₂) also gives a characteristic Amide II band
    (N-H bend + C-N stretch) at ~1620 cm⁻¹, which overlaps with triazole peaks.
    This overlap is normal for ribavirin and accounted for in the reference spectrum.
    Ribofuranosyl acetate impurity (protected synthesis precursor) manifests as
    additional ester C=O peaks at ~1740 cm⁻¹. Presence of a peak in this region
    requires quantitative impurity control by HPLC.

input.csv

BatchNumber,CorrelationCoefficient,Peak3400_cm1,Peak1680_cm1,Peak1620_cm1,Peak1210_cm1,Peak1050_cm1,BaselineNoiseAbs,PreparationMethod
RBV-ID-2026-027,0.9994,3395.0,1678.0,1622.0,1208.0,1052.0,0.0014,KBr
RBV-ID-2026-028,0.9908,3430.0,1665.0,1635.0,1195.0,1068.0,0.007,KBr

URS & FS — User Requirements and Functional Specification

This document defines the controlled interface, validation behaviour and expected output of InfraLUM_ID_Ribavirin. The document is intended to reside inside the utility folder together with input.csv and the description file. Expected executable: InfraLUM_ID_Ribavirin.exe.

Purpose: the utility performs deterministic verification of QC input data against approved limits, generates machine-readable output.json, and supports QC/QA review, LIMS/ELN/MES integration and CSV/CSA preparation.

Scope

Utility classIR spectroscopy / InfraLUM
Main object/methodRibavirin Identification by IR Spectroscopy
Inputinput.csv with a fixed header and controlled data types.
Outputoutput.json with PASS, WARNING, FAIL statuses, detailed checks and critical findings.
ExclusionsThe utility does not replace the approved specification, analytical procedure, OOS/OOT investigation or Qualified Person / QA release decision.

Domain limits and critical parameters

Key fragments from the source description are listed below. Before production use, limits shall be verified against the approved specification, registration dossier, current pharmacopoeial edition and local SOPs.
  • • Correlation coefficient with reference spectrum: ≥0.995
  • • Peak 3400 cm⁻¹ (O-H/N-H stretching, hydrogen bonds): ±25 cm⁻¹
  • • Peak 1680 cm⁻¹ (carboxamide C=O): ±10 cm⁻¹ — AMIDE MARKER
  • • Peak 1620 cm⁻¹ (triazole C=N/NH₂): ±10 cm⁻¹
  • • Peak 1210 cm⁻¹ (ribose C-O-C): ±10 cm⁻¹ — SUGAR MARKER
  • • Peak 1050 cm⁻¹ (primary alcohol C-O CH₂OH): ±10 cm⁻¹
  • • Baseline noise: ≤0.005 Abs
  • ⚠️ IMPORTANT: Ribavirin is a synthetic guanosine nucleoside analogue.
  • creating a very broad absorption band at 3200–3500 cm⁻¹.
  • InfraLUM_ID_Ribavirin.exe input.csv output.json → evaluate your data
  • BatchNumber,CorrelationCoefficient,Peak3400_cm1,Peak1680_cm1,Peak1620_cm1,Peak1210_cm1,Peak1050_cm1,BaselineNoiseAbs,PreparationMethod
  • • Used in combination with interferon or sofosbuvir for hepatitis C treatment.
  • • Mechanism: viral RNA polymerase inhibition and intracellular GTP pool depletion.
  • • IR spectroscopy is the primary identity confirmation method per Ph. Eur. and Russian Pharmacopoeia.
  • • Reference spectrum established in Ph. Eur. Ribavirin and Russian Pharmacopoeia FS.2.4.00270.18.
  • • InfraLUM FT-08/FT-10 instruments are widely used in Russian QC laboratories.

Pharmacopoeial and regulatory references found in the description

#SourceNote
1Ph. EurUsed as a source/context reference from the description; the current edition must be verified before implementation.
2Ph. Eur. RibavirinUsed as a source/context reference from the description; the current edition must be verified before implementation.
3Ph. Eur. / Russian PharmacopoeiaUsed as a source/context reference from the description; the current edition must be verified before implementation.
4Ph. Eur. and Russian PharmacopoeiaUsed as a source/context reference from the description; the current edition must be verified before implementation.
5Ph. Eur. Ribavirin and Russian Pharmacopoeia FS.2.4.00270.18Used as a source/context reference from the description; the current edition must be verified before implementation.
6uspectedUsed as a source/context reference from the description; the current edition must be verified before implementation.
7ГФ РФ / Ph. EurUsed as a source/context reference from the description; the current edition must be verified before implementation.
8ГФ РФ и Ph. EurUsed as a source/context reference from the description; the current edition must be verified before implementation.
9ГФ РФ ФС.2.4.00270.18 и Ph. Eur. RibavirinUsed as a source/context reference from the description; the current edition must be verified before implementation.

Production data flow

StepResponsible objectDescriptionControl
1LIMS/ELN/MES or QC analystCreates input.csv with measured values and batch identifier.Header, mandatory-field and unit checks.
2InfraLUM_ID_RibavirinLoads CSV, converts data types, applies domain rules and produces the result.Deterministic rules; no ML/probabilistic conformance decision.
3QC reviewer / QAReviews output.json, warnings, critical deviations and source values.Review, deviation investigation and data-trail retention.
4Archive / eQMSRetains input, output, utility version, documentation and checksum.Data integrity: ALCOA+, reproducibility and change control.

URS — User Requirements Specification

IDRequirementCriticalityAcceptance criterion
URS-001The utility shall be supplied with InfraLUM_ID_Ribavirin.documentation.html in the same folder as the description and sample input.csv.MediumThe documentation file exists in the InfraLUM_ID_Ribavirin folder and opens locally in a browser.
URS-002The utility shall accept input.csv with the exact headers defined in the data contract.HighA CSV file with the correct header is processed without manual editing.
URS-003The utility shall validate mandatory fields, column count, data types and basic plausibility of values.HighSchema and conversion errors are explicitly reported.
URS-004The utility shall evaluate each critical and non-critical QC parameter against approved limits.HighEach checked parameter receives a status and explanatory message.
URS-005The utility shall produce a final record/batch status: PASS, WARNING or FAIL.HighA critical deviation causes FAIL; a non-critical deviation causes WARNING.
URS-006The utility shall generate output.json suitable for LIMS/ELN/MES integration and QA/QC review.HighJSON contains source values, rules, statuses, warnings and critical findings.
URS-007The utility shall preserve traceability between batch identifier, input values, applied rules and final status.HighThe output contains the batch/row identifier and list of checked parameters.
URS-008The utility shall not use machine learning or probabilistic interpretation for conformance decisions.MediumThe result is reproducible and based on explicit rules, thresholds and input values.
URS-009The utility shall support the CLI scenario: InfraLUM_ID_Ribavirin.exe input.csv output.json.HighWith valid arguments, output.json is created; on error, a diagnostic message is returned.
URS-010The documentation shall support IQ/OQ/PQ or equivalent CSV/CSA verification.MediumThe document contains URS, FS, CSV contract, rules, test scenarios and traceability.
URS-011The utility shall distinguish data errors, method warnings and critical specification failures.Highoutput.json provides separate structures/fields for errors, warnings and critical findings.
URS-012The utility shall support archival retention of unchanged input and output data.Mediuminput.csv and output.json can be retained as a batch record or OQ/PQ evidence.

input.csv contract

#FieldTypeSamplePurpose
1BatchNumberstringRBV-ID-2026-027Batch or lot identifier used for traceability.
2CorrelationCoefficientdecimal0.9994Similarity/correlation criterion against a reference profile.
3Peak3400_cm1decimal3395.0Spectral peak position or chromatographic/analytical peak attribute.
4Peak1680_cm1decimal1678.0Spectral peak position or chromatographic/analytical peak attribute.
5Peak1620_cm1decimal1622.0Spectral peak position or chromatographic/analytical peak attribute.
6Peak1210_cm1decimal1208.0Spectral peak position or chromatographic/analytical peak attribute.
7Peak1050_cm1decimal1052.0Spectral peak position or chromatographic/analytical peak attribute.
8BaselineNoiseAbsdecimal0.0014Signal quality parameter used to determine analytical suitability.
9PreparationMethodcontrolled stringKBrControlled analytical or sample-preparation method identifier.
BatchNumber,CorrelationCoefficient,Peak3400_cm1,Peak1680_cm1,Peak1620_cm1,Peak1210_cm1,Peak1050_cm1,BaselineNoiseAbs,PreparationMethod
RBV-ID-2026-027,0.9994,3395.0,1678.0,1622.0,1208.0,1052.0,0.0014,KBr
RBV-ID-2026-028,0.9908,3430.0,1665.0,1635.0,1195.0,1068.0,0.007,KBr

Input validation rules

IDFieldTypeCheckCriticality
VR-001BatchNumberstringnot empty; unique within file where applicableMedium
VR-002CorrelationCoefficientdecimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-003Peak3400_cm1decimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-004Peak1680_cm1decimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-005Peak1620_cm1decimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-006Peak1210_cm1decimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-007Peak1050_cm1decimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-008BaselineNoiseAbsdecimalparse as invariant decimal; finite value; no NaN/Infinity; evaluate against approved numerical limitMedium
VR-009PreparationMethodcontrolled stringnot empty; value belongs to controlled set or is reported as warningMedium

FS — Functional Specification

IDFunctionImplementationURS link
FS-001Local HTML documentationProvide a bilingual InfraLUM_ID_Ribavirin.documentation.html file with RU/EN switch, source description, URS, FS, contracts and tests.URS-001, URS-010
FS-002CSV importRead input.csv in UTF-8/CSV-compatible format and identify the header row and data records.URS-002, URS-009
FS-003Schema validationCompare actual columns against the expected contract; missing key fields are recorded as schema errors.URS-002, URS-003
FS-004Type conversionConvert decimal, flag/boolean and controlled string values. Invalid values are recorded at row and field level.URS-003
FS-005Domain rule engineApply rules for InfraLUM ID Ribavirin, including critical limits from the description and approved specification.URS-004, URS-008
FS-006Status aggregationAggregate parameter statuses: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance.URS-005, URS-011
FS-007JSON exportWrite output.json with utility identifier, source values, check results, warnings and critical findings.URS-006, URS-007
FS-008Error handlingSeparate execution errors, schema errors, conversion errors and domain-limit failures.URS-003, URS-011
FS-009Audit supportKeep result structure suitable for review, OOS/OOT investigation, calculation reproduction and inspection discussion.URS-007, URS-012
FS-010Integration contractSupport the standard scenario: LIMS/ELN/MES creates input.csv, the utility returns output.json, and the portal displays description and documentation.URS-006, URS-009
FS-011Configuration/change controlAny change to limits, column names or rule logic requires versioning, review, regression tests and documentation update.URS-010, URS-012
FS-012Data integrityDo not modify the source input.csv; save results separately in output.json for archival retention.URS-007, URS-012

Example output.json structure

{
  "utilityId": "infralum-id-ribavirin",
  "utilityName": "InfraLUM_ID_Ribavirin",
  "object": "InfraLUM ID Ribavirin",
  "overallStatus": "PASS|WARNING|FAIL",
  "sourceFile": "input.csv",
  "recordCount": 2,
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "RBV-ID-2026-027",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-001"
    },
    {
      "parameter": "CorrelationCoefficient",
      "value": "0.9994",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-002"
    },
    {
      "parameter": "Peak3400_cm1",
      "value": "3395.0",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-003"
    },
    {
      "parameter": "Peak1680_cm1",
      "value": "1678.0",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-004"
    },
    {
      "parameter": "Peak1620_cm1",
      "value": "1622.0",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-005"
    },
    {
      "parameter": "Peak1210_cm1",
      "value": "1208.0",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-006"
    },
    {
      "parameter": "Peak1050_cm1",
      "value": "1052.0",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-007"
    },
    {
      "parameter": "BaselineNoiseAbs",
      "value": "0.0014",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-008"
    },
    {
      "parameter": "PreparationMethod",
      "value": "KBr",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result",
      "ruleReference": "VR-009"
    }
  ],
  "criticalFindings": [],
  "warnings": [],
  "metadata": {
    "executionMode": "CLI",
    "interface": "CSV to JSON",
    "documentationFile": "InfraLUM_ID_Ribavirin.documentation.html"
  }
}

Traceability matrix URS → FS → OQ/PQ

URSFSVerificationExpected result
URS-001FS-001Check that InfraLUM_ID_Ribavirin.documentation.html exists in the utility folder.The file exists, opens, and the RU/EN switch works.
URS-002/003FS-002/003/004Run the utility with a correct CSV, a CSV missing a mandatory column and a CSV with an invalid data type.Correct CSV is processed; schema and type errors are explicit.
URS-004/005FS-005/006Submit values within limits, at warning level and beyond a critical limit.PASS, WARNING and FAIL are produced according to rules.
URS-006/007FS-007/009/012Verify output.json structure and batch traceability.JSON contains batch identifier, values, rules, statuses and critical findings.
URS-008FS-005Repeat execution with the same input.csv.The result is identical; no probabilistic behaviour is present.
URS-009/010/012FS-010/011/012Execute the CLI scenario and archive input/output/version.Artefacts are suitable for OQ/PQ and change-control records.

OQ/PQ test scenarios

IDScenarioInput dataExpected result
OQ-001Positive pathSample CSV with valid values.output.json is created; final status is PASS or acceptable WARNING according to rules.
OQ-002Mandatory column missingRemove one key column from the header.Schema error; the row shall not be treated as passed.
OQ-003Invalid typePut text into a numeric field.Type-conversion error with field and row identification.
OQ-004Critical limitCritical parameter value outside the approved limit.FAIL and an entry in criticalFindings.
OQ-005WarningNon-critical deviation or unknown controlled category.WARNING without hiding the source value.
PQ-001User real batchReal input.csv from the QC process.Reviewed result with input.csv, output.json, version and checksum retained.

QA/QC, CSV/CSA and change control

  • Do not rename columns without updating the validator, test set and documentation.
  • Retain input.csv, output.json, executable version, checksum and a copy of the documentation.
  • Before production use, perform IQ/OQ/PQ or equivalent risk-based CSV/CSA verification.
  • Critical limits shall be approved by the process owner and verified against the registration dossier, pharmacopoeia and local SOPs.
  • Any change to rules, thresholds or status interpretation shall go through change control and regression testing.

Included in packages

Lumex QC Suite

A single Lumex package combining the former Lumex and Lumex2 sets: instrumental and general pharmaceutical QC, AAS/ICP, CE, HPLC, NIR/PAT, stability, dissolution, content uniformity, system suitability and statistical control.