HyaluronicAcidFillerQualityChecker

Hyaluronic Acid Filler

dermal fillers endotoxins hyaluronic acid injectables particles rheology sterility
Open selection

HA Filler Quality Checker — Hyaluronic Acid Dermal Fillers (ISO 13485, EU MDR)

ℹ️ The utility checks HA concentration, crosslinking degree, residual BDDE, rheology (G', viscosity), particle size, osmolality, sterility, endotoxins.
ℹ️ Does not use machine learning — rule-based logic only.
⚠️ CRITICAL! Residual BDDE ≤ 2 ppm (toxic). Sterility and endotoxins mandatory. Rheology determines injection site!

Usage:
HyaluronicAcidFillerQualityChecker.exe → demo mode
HyaluronicAcidFillerQualityChecker.exe input.csv output.json → evaluate data

Input format:
BatchNumber,HA_Concentration_mgPerMl,Crosslinking_Degree_Percent,Residual_BDDE_ppm,Particle_Size_um,Viscosity_Pa_s,G_Prime_Pa,G_DoublePrime_Pa,Swelling_Ratio,pH_Value,Osmolality_mOsm_kg,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Protein_Residue_ppm

Example:
HA-FILL-2026-001,24.0,1.5,1.2,250.0,25.0,450.0,150.0,12.0,7.1,300.0,0.15,0,0,5.0

— WHY IS THIS NEEDED?
Hyaluronic Acid (HA) dermal fillers — most popular injectable products for contour plasty:
• Juvederm, Restylane, Teosyal, Belotero, Revanesse etc.
• Crosslinked HA provides long-term volume and collagen stimulation
• Rheological properties (G', viscosity) dictate where to inject (deep, superficial, lips, tear trough)
⚠️ CRITICAL:
• Residual BDDE ≤ 2 ppm — 1,4-butanediol diglycidyl ether used for crosslinking, but TOXIC in free form. Strict limit!
• Crosslinking Degree 0.5–4.0% — determines gel density and degradation rate
• HA Concentration 20–26 mg/mL — standard for most medium-density fillers
• Elastic Modulus (G') — key selection parameter: high G' for bone/cartilage, low G' for superficial wrinkles
• Particle Size — must match needle gauge (typically < 500 µm for 27G/30G passage)
• Osmolality 250–350 mOsm/kg — isotonicity prevents post-injection swelling
• Sterility and Endotoxins — MANDATORY for all injectable forms
• Protein (for non-biofermented HA) — ≤ 10 ppm (allergy risk)
⚠️ Note: Rheology (G', G'', tan δ) measured on rheometer. G' (elastic modulus) shows ability to restore shape (lifting capacity). High G' = high lifting force. Residual BDDE removed by multi-step purification, its control critical for safety. Osmolality must be close to physiological to prevent fluid influx and swelling. Sterility ensured by terminal sterilization or aseptic filling.

input.csv

BatchNumber,HA_Concentration_mgPerMl,Crosslinking_Degree_Percent,Residual_BDDE_ppm,Particle_Size_um,Viscosity_Pa_s,G_Prime_Pa,G_DoublePrime_Pa,Swelling_Ratio,pH_Value,Osmolality_mOsm_kg,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Protein_Residue_ppm
HA-FILL-2026-001,24.0,1.5,1.2,250.0,25.0,450.0,150.0,12.0,7.1,300.0,0.15,0,0,5.0

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: HyaluronicAcidFillerQualityCheckerAPI / object: Hyaluronic Acid FillerCategory: Dermal fillers

1. Scope

The Hyaluronic Acid Filler Quality Checker utility is used for Hyaluronic Acid Filler. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Dermal fillers

Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility

2. Execution modes

HyaluronicAcidFillerQualityChecker.exe                            → demo mode
HyaluronicAcidFillerQualityChecker.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • assay / content
  • microbiology and pathogens
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ CRITICAL! Residual BDDE ≤ 2 ppm (toxic). Sterility and endotoxins mandatory. Rheology determines injection site!
  • ⚠️ CRITICAL:
  • Residual BDDE ≤ 2 ppm — 1,4-butanediol diglycidyl ether used for crosslinking, but TOXIC in free form. Strict limit!
  • Crosslinking Degree 0.5–4.0% — determines gel density and degradation rate
  • HA Concentration 20–26 mg/mL — standard for most medium-density fillers
  • Elastic Modulus (G') — key selection parameter: high G' for bone/cartilage, low G' for superficial wrinkles
  • Particle Size — must match needle gauge (typically < 500 µm for 27G/30G passage)
  • Osmolality 250–350 mOsm/kg — isotonicity prevents post-injection swelling
  • Sterility and Endotoxins — MANDATORY for all injectable forms
  • Protein (for non-biofermented HA) — ≤ 10 ppm (allergy risk)
  • ⚠️ Note: Rheology (G', G'', tan δ) measured on rheometer. G' (elastic modulus) shows ability to restore shape (lifting capacity). High G' = high lifting force. Residual BDDE removed by multi-step purification, its control critical for safety. Osmolality must be close to physiological to prevent fluid influx and swelling. Sterility ensured by terminal sterilization or aseptic filling.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Hyaluronic Acid Filler with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content, microbiology and pathogens, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberHA-FILL-2026-001mandatory field; used for batch/lot traceability
2HA_Concentration_mgPerMldecimalmg/mlHA Concentration mg per Ml24.0numeric value; compared with assay/purity limit from specification or method
3Crosslinking_Degree_Percentdecimal%Crosslinking Degree %1.5mandatory numeric value; rule comparison is performed by the utility
4Residual_BDDE_ppmdecimalppmResidual BDDE ppm1.2mandatory numeric value; rule comparison is performed by the utility
5Particle_Size_umdecimalµmParticle Size um250.0mandatory numeric value; rule comparison is performed by the utility
6Viscosity_Pa_sdecimalas specifiedViscosity Pa s25.0mandatory numeric value; rule comparison is performed by the utility
7G_Prime_Padecimalas specifiedG Prime Pa450.0mandatory numeric value; rule comparison is performed by the utility
8G_DoublePrime_Padecimalas specifiedG Double Prime Pa150.0mandatory numeric value; rule comparison is performed by the utility
9Swelling_Ratiodecimalas specifiedSwelling Ratio12.0mandatory numeric value; rule comparison is performed by the utility
10pH_ValuedecimalpHp H Value7.1mandatory numeric value; rule comparison is performed by the utility
11Osmolality_mOsm_kgdecimalmOsm/kgOsmolality m Osm kg300.0mandatory numeric value; rule comparison is performed by the utility
12Endotoxin_EU_mldecimalas specifiedEndotoxin EU ml0.15numeric value; critical safety attribute compared with endotoxin limit
13SterilityTestboolean / flag0/1Sterility Test0valid flag required; prohibited organisms and growth are normally expected as 0 / absent
14PackagingIntegrityTestdecimalas specifiedPackaging Integrity Test0mandatory numeric value; rule comparison is performed by the utility
15Protein_Residue_ppmdecimalppmProtein Residue ppm5.0mandatory numeric value; rule comparison is performed by the utility

CSV example

BatchNumber,HA_Concentration_mgPerMl,Crosslinking_Degree_Percent,Residual_BDDE_ppm,Particle_Size_um,Viscosity_Pa_s,G_Prime_Pa,G_DoublePrime_Pa,Swelling_Ratio,pH_Value,Osmolality_mOsm_kg,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Protein_Residue_ppm
HA-FILL-2026-001,24.0,1.5,1.2,250.0,25.0,450.0,150.0,12.0,7.1,300.0,0.15,0,0,5.0

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable HyaluronicAcidFillerQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Hyaluronic Acid Filler, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "HyaluronicAcidFillerQualityChecker",
  "api": "Hyaluronic Acid Filler",
  "batchNumber": "HA-FILL-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "HA-FILL-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "HA_Concentration_mgPerMl",
      "value": "24.0",
      "unit": "mg/ml",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Crosslinking_Degree_Percent",
      "value": "1.5",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Residual_BDDE_ppm",
      "value": "1.2",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Particle_Size_um",
      "value": "250.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Viscosity_Pa_s",
      "value": "25.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "G_Prime_Pa",
      "value": "450.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "G_DoublePrime_Pa",
      "value": "150.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-005Positive pathogen/microbiological flag or microbiological limit excursionCritical FAIL is produced with the offending parameter.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Dermal Fillers QC Suite

Package for injectable implants and dermal fillers: hyaluronic acid, CaHA, PLLA, PMMA, collagen and related tests.