HelioxMixtureQualityChecker

Heliox Mixture

carbon dioxide impurities medical gases nitrogen oxygen specification water quality
Open selection

Heliox Mixture Quality Checker — Medical heliox mixture quality assessment (Ph. Eur. 0439)

ℹ️ Verifies compliance with European Pharmacopoeia for heliox (typically 80% He / 20% O₂).
⚠️ Critical limits: He = 80.0±1.0%, CO ≤5 ppm, H₂O ≤67 ppm.

Usage:
HelioxMixtureQualityChecker.exe → demo mode
HelioxMixtureQualityChecker.exe input.csv output.json → evaluate data

Input format:
ProductName,BatchNumber,Helium_percent,Oxygen_percent,CarbonMonoxide_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance

Example:
Heliox_80_20_Medical,G20250306,80.1,19.9,0.2,40,0.05,"ColorlessOdorlessGas"

— WHY IS THIS NEEDED?
Heliox is used in severe asthma, COPD, and pediatric croup. Composition accuracy is critical:
• Excess O₂ reduces gas density, diminishing therapeutic effect,
• Insufficient O₂ causes hypoxia,
• CO impurity is toxic even at trace levels.
Compliance with Ph. Eur. 0439 and internal specification is mandatory for batch release.

input.csv

ProductName,BatchNumber,Helium_percent,Oxygen_percent,CarbonMonoxide_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance
Heliox_80_20_Medical,G20250306,80.1,19.9,0.2,40,0.05,"ColorlessOdorlessGas"

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: HelioxMixtureQualityCheckerAPI / object: Heliox MixtureCategory: Medical gases

1. Scope

The Heliox Mixture Quality Checker utility is used for Heliox Mixture. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Medical gases

Tags: carbon dioxide, impurities, medical gases, nitrogen, oxygen, specification, water quality

2. Execution modes

HelioxMixtureQualityChecker.exe                            → demo mode
HelioxMixtureQualityChecker.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • assay / content

4. Domain limits and critical parameters

  • ⚠️ Critical limits: He = 80.0±1.0%, CO ≤5 ppm, H₂O ≤67 ppm.
  • Heliox is used in severe asthma, COPD, and pediatric croup. Composition accuracy is critical:
  • Excess O₂ reduces gas density, diminishing therapeutic effect,
  • Insufficient O₂ causes hypoxia,
  • CO impurity is toxic even at trace levels.
  • Compliance with Ph. Eur. 0439 and internal specification is mandatory for batch release.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Heliox Mixture with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1ProductNameidentifieras specifiedProduct NameHeliox_80_20_Medicalmandatory value; format and acceptability are checked by the utility
2BatchNumberidentifieras specifiedBatch NumberG20250306mandatory field; used for batch/lot traceability
3Helium_percentdecimal%Helium percent80.1mandatory numeric value; rule comparison is performed by the utility
4Oxygen_percentdecimal%Oxygen percent19.9mandatory numeric value; rule comparison is performed by the utility
5CarbonMonoxide_ppmdecimalppmCarbon Monoxide ppm0.2mandatory numeric value; rule comparison is performed by the utility
6WaterContent_ppmdecimalppmWater Content ppm40numeric value; compared with assay/purity limit from specification or method
7TotalHydrocarbons_as_methane_ppmdecimalppmTotal Hydrocarbons as methane ppm0.05mandatory numeric value; rule comparison is performed by the utility
8Appearancetextas specifiedAppearanceColorlessOdorlessGasmandatory value; format and acceptability are checked by the utility

CSV example

ProductName,BatchNumber,Helium_percent,Oxygen_percent,CarbonMonoxide_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance
Heliox_80_20_Medical,G20250306,80.1,19.9,0.2,40,0.05,"ColorlessOdorlessGas"

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable HelioxMixtureQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Heliox Mixture, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "HelioxMixtureQualityChecker",
  "api": "Heliox Mixture",
  "batchNumber": "G20250306",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "ProductName",
      "value": "Heliox_80_20_Medical",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "BatchNumber",
      "value": "G20250306",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Helium_percent",
      "value": "80.1",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Oxygen_percent",
      "value": "19.9",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "CarbonMonoxide_ppm",
      "value": "0.2",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "WaterContent_ppm",
      "value": "40",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "TotalHydrocarbons_as_methane_ppm",
      "value": "0.05",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Appearance",
      "value": "ColorlessOdorlessGas",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Medical Gases QC Suite

Quality-control package for medical and pharmaceutical gases: oxygen, nitrogen, carbon dioxide, nitrous oxide, helium and mixtures.

Open