DisintegrationTabletCapsuleQualityChecker

Disintegration Tablet/Capsule

LabEx laboratory QC CSV→JSON URS & FS solid dosage forms utilities / cleanroom
Open selection
Disintegration Tablet/Capsule Quality Checker — Control of Tablet and Capsule Disintegration

ℹ️  Utility checks compliance of disintegration time with USP <701> / Ph. Eur. 2.9.1 requirements:
   • Stage 1: 6 units tested simultaneously.
   • Passing Criteria Stage 1: All 6 units must disintegrate within the specified time.
   • If 1 or 2 units fail, Stage 2 is conducted (additional 12 units).
   • Standard limits: 15 minutes (uncoated tablets), 30 minutes (capsules).

⚠️  CRITICAL: Presence of residue on the apparatus screen is considered non-disintegration!
   Exceeding disintegration time may indicate issues with granulation or compression.

Usage:
 DisintegrationTabletCapsuleQualityChecker.exe                            → demo mode (console output)
 DisintegrationTabletCapsuleQualityChecker.exe input.csv output.json      → evaluate your data

Input format:
BatchNumber,LimitMinutes,Time1,Time2,Time3,Time4,Time5,Time6

Example:
 DIS-TAB-2026-001,15.0,8.5,9.0,8.0,9.5,8.2,9.1

— WHY IS THIS NEEDED?
Disintegration is a critical quality attribute for immediate-release solid dosage forms.
• Ensures availability of the drug substance for subsequent dissolution and absorption.
• USP <701> and Ph. Eur. 2.9.1 regulate the test method using basket-rack assemblies.
• The test is faster and simpler than full Dissolution testing and is often used for in-process control.
• Automation of calculation and criteria check eliminates interpretation errors of Stage 1/Stage 2 results.

⚠️  CRITICAL:
• Medium temperature must be maintained at 37 ± 2 °C.
• For coated tablets, disk assemblies may be used.
• If at least one unit fails Stage 1, the result is considered unsatisfactory until Stage 2 is performed.
• Stage 2 is passed if not less than 16 of 18 tested units (6+12) meet the limit.

Key features:
• Automatic check of Stage 1 criteria.
• Calculation of maximum and mean disintegration time.
• Generation of LIMS-compatible JSON report.

Critical parameters:
• Max Time: <= Limit (e.g., 15 min)
• Failed Units (Stage 1): 0

💡 Usage tips:
1. Ensure the basket rack moves freely up and down.
2. Use degassed water or buffer solution if specified in the monograph.
3. For gelatin capsules, preliminary soaking in water may be required.
4. Visually confirm complete disappearance of residue on the screen.
5. Regularly calibrate time and temperature of the disintegration apparatus.

⚠️ Note: Disintegration does not always correlate with Dissolution. For modified-release products or poorly soluble drugs, the disintegration test may be replaced or supplemented by the dissolution test.

input.csv

BatchNumber,LimitMin,Time1,Time2,Time3,Time4,Time5,Time6
DIS-TAB-2026-001,15.0,8.5,9.0,8.0,9.5,8.2,9.1
DIS-CAP-2026-001,30.0,12.0,14.5,13.0,15.0,12.5,14.0
DIS-FAIL-2026-001,15.0,10.0,12.0,18.0,11.0,13.0,14.0

URS & FS — Disintegration Tablet/Capsule

This document describes the user requirements and functional specification for DisintegrationTabletCapsuleQualityChecker. The utility is intended for QC laboratory use as a rule-based check of data prepared from an instrument, LIMS, ELN, MES or an approved input.csv.

URS — User Requirements Specification

IDRequirementCriticalityAcceptance criterion
URS-001The utility shall accept input.csv with approved columns: BatchNumber, LimitMin, Time1, Time2, Time3, Time4, Time5, Time6.HighThe CSV file is processed without manual header editing.
URS-002The utility shall perform deterministic evaluation for “Disintegration Tablet/Capsule” without machine learning.HighThe same input data produce the same JSON result.
URS-003The utility shall validate mandatory fields, numeric formats, flags, ranges and domain plausibility.HighSchema and conversion errors are explicitly reported.
URS-004The utility shall generate output.json with PASS / WARNING / FAIL statuses, source values, warnings and failures.HighThe JSON result is suitable for review, deviation investigation and integration.
URS-005The documentation shall support IQ/OQ/PQ or CSV/CSA verification.MediumURS/FS, CSV/JSON contract and test scenarios are supplied with the utility.

input.csv contract

#FieldSamplePurpose
1BatchNumberDIS-TAB-2026-001Input parameter for deterministic QC evaluation.
2LimitMin15.0Input parameter for deterministic QC evaluation.
3Time18.5Input parameter for deterministic QC evaluation.
4Time29.0Input parameter for deterministic QC evaluation.
5Time38.0Input parameter for deterministic QC evaluation.
6Time49.5Input parameter for deterministic QC evaluation.
7Time58.2Input parameter for deterministic QC evaluation.
8Time69.1Input parameter for deterministic QC evaluation.

FS — Functional Specification

IDFunctionImplementation
FS-001CSV importRead input.csv; validate header, column count and encoding.
FS-002Schema validationCheck required fields and permitted input values.
FS-003Rule engineApply domain rules, limits and system suitability/specification checks described in the source utility description.
FS-004Status aggregationProduce final status: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance.
FS-005JSON exportWrite machine-readable output.json for LIMS/ELN/MES and QA/QC review.

OQ/PQ scenarios

  • OQ-001: a valid sample row shall be processed without schema error.
  • OQ-002: a missing mandatory column shall produce a schema error.
  • OQ-003: a non-numeric value in a numeric field shall produce a conversion error.
  • OQ-004: a critical parameter outside the limit shall produce FAIL or a critical finding.
  • PQ-001: user real batches shall be checked with retention of input.csv, output.json, utility version and checksum.

Included in packages

Dental & Dental Materials QC Suite

QC utility package for dental medicinal products, oral-care products and dental materials: anaesthetics, antiseptics, analgesic/anti-infective products, polymers/minerals, resins, bonding workflows, microbiology, stability and material QC.

Open

LabEx QC Suite

LabEx package for laboratory QC methods: chromatography, spectroscopy, microbiology, sterility, endotoxins, particles, dissolution, water, cleanrooms and sample preparation.

Open