CarbonDioxidePharmaceuticalQualityChecker

Carbon Dioxide Pharmaceutical

carbon dioxide impurities medical gases nitrogen oxygen specification water quality
Open selection

Carbon Dioxide Pharmaceutical Quality Checker — Pharmaceutical CO₂ quality assessment (Ph. Eur. 0423)

ℹ️ Verifies compliance with European Pharmacopoeia.
⚠️ Critical limits: CO ≤10 ppm, O₂ ≤500 ppm, H₂O ≤67 ppm.

Usage:
CarbonDioxidePharmaceuticalQualityChecker.exe → demo mode
CarbonDioxidePharmaceuticalQualityChecker.exe input.csv output.json → evaluate data

Input format:
ProductName,BatchNumber,Purity_percent,CarbonMonoxide_ppm,Oxygen_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance

Example:
CarbonDioxide_Pharmaceutical_Gas,G20250304,99.9,5.0,300,50,0.1,"ColorlessOdorlessGas"

— WHY IS THIS NEEDED?
Pharmaceutical CO₂ is used in laparoscopy (pneumoperitoneum), respiratory therapy, and as a carbogen component. CO impurity is toxic, excess O₂ disrupts gas balance, and moisture causes equipment corrosion.
Compliance with Ph. Eur. 0423 is mandatory for batch release.

input.csv

ProductName,BatchNumber,Purity_percent,CarbonMonoxide_ppm,Oxygen_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance
CarbonDioxide_Pharmaceutical_Gas,G20250304,99.9,5.0,300,50,0.1,"ColorlessOdorlessGas"

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: CarbonDioxidePharmaceuticalQualityCheckerAPI / object: Carbon Dioxide PharmaceuticalCategory: Medical gases

1. Scope

The Carbon Dioxide Pharmaceutical Quality Checker utility is used for Carbon Dioxide Pharmaceutical. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Medical gases

Tags: carbon dioxide, impurities, medical gases, nitrogen, oxygen, specification, water quality

2. Execution modes

CarbonDioxidePharmaceuticalQualityChecker.exe                            → demo mode
CarbonDioxidePharmaceuticalQualityChecker.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • assay / content
  • purity and impurities

4. Domain limits and critical parameters

  • ⚠️ Critical limits: CO ≤10 ppm, O₂ ≤500 ppm, H₂O ≤67 ppm.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Carbon Dioxide Pharmaceutical with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content, purity and impurities.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1ProductNameidentifieras specifiedProduct NameCarbonDioxide_Pharmaceutical_Gasmandatory value; format and acceptability are checked by the utility
2BatchNumberidentifieras specifiedBatch NumberG20250304mandatory field; used for batch/lot traceability
3Purity_percentdecimal%Purity percent99.9numeric value; compared with assay/purity limit from specification or method
4CarbonMonoxide_ppmdecimalppmCarbon Monoxide ppm5.0mandatory numeric value; rule comparison is performed by the utility
5Oxygen_ppmdecimalppmOxygen ppm300mandatory numeric value; rule comparison is performed by the utility
6WaterContent_ppmdecimalppmWater Content ppm50numeric value; compared with assay/purity limit from specification or method
7TotalHydrocarbons_as_methane_ppmdecimalppmTotal Hydrocarbons as methane ppm0.1mandatory numeric value; rule comparison is performed by the utility
8Appearancetextas specifiedAppearanceColorlessOdorlessGasmandatory value; format and acceptability are checked by the utility

CSV example

ProductName,BatchNumber,Purity_percent,CarbonMonoxide_ppm,Oxygen_ppm,WaterContent_ppm,TotalHydrocarbons_as_methane_ppm,Appearance
CarbonDioxide_Pharmaceutical_Gas,G20250304,99.9,5.0,300,50,0.1,"ColorlessOdorlessGas"

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable CarbonDioxidePharmaceuticalQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Carbon Dioxide Pharmaceutical, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "CarbonDioxidePharmaceuticalQualityChecker",
  "api": "Carbon Dioxide Pharmaceutical",
  "batchNumber": "G20250304",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "ProductName",
      "value": "CarbonDioxide_Pharmaceutical_Gas",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "BatchNumber",
      "value": "G20250304",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Purity_percent",
      "value": "99.9",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "CarbonMonoxide_ppm",
      "value": "5.0",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Oxygen_ppm",
      "value": "300",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "WaterContent_ppm",
      "value": "50",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "TotalHydrocarbons_as_methane_ppm",
      "value": "0.1",
      "unit": "ppm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Appearance",
      "value": "ColorlessOdorlessGas",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Medical Gases QC Suite

Quality-control package for medical and pharmaceutical gases: oxygen, nitrogen, carbon dioxide, nitrous oxide, helium and mixtures.

Open