TOC Water Quality Checker — Control of Total Organic Carbon in Water
ℹ️ Utility checks the quality of Purified Water (PW) and Water for Injection (WFI) according to USP <643> (TOC) and USP <645> (Conductivity):
• Total Organic Carbon (TOC): ≤ 500 ppb (µg/L)
• Conductivity: ≤ 1.3 µS/cm at 25°C (Stage 1)
• System Suitability (SST): Standard response 85-115% of theoretical
• Blank: ≤ 100 ppb (background control)
⚠️ CRITICAL: High TOC level (>500 ppb) indicates possible bacterial contamination, resin degradation, or organic leakage!
Conductivity exceedance indicates insufficient demineralization.
Usage:
TOCWaterQualityChecker.exe → demo mode (console output)
TOCWaterQualityChecker.exe input.csv output.json → evaluate your data
Input format:
BatchNumber,WaterType,TOC_ppb,Conductivity_uS_cm,Temperature_C,ResponseStandard_ppb,BlankResponse_ppb
Example:
WFI-2026-LOOP-001,WFI,120.0,0.9,25.0,480.0,50.0
— WHY IS THIS NEEDED?
Water is the most used component in pharmaceutical manufacturing.
TOC control replaces the obsolete "Oxidizable Substances" test and is a more sensitive indicator of organic contamination.
Conductivity controls the presence of ionic impurities. These two parameters are key for water release into production.
⚠️ CRITICAL:
• TOC ≤ 500 ppb — strict limit for PW and WFI worldwide (USP, EP, JP).
• Conductivity depends on temperature. At 25°C limit is 1.3 µS/cm. For other temperatures, USP <645> tables are used.
• System Suitability (SST) must be performed daily or before a series of measurements using sucrose or benzoquinone standard.
• High background (Blank) makes measurements unreliable.
Key features:
• Simultaneous check of TOC and Conductivity
• System Suitability (SST) control
• Temperature consideration for conductivity interpretation
• Generation of reports for Engineering and QC departments, and LIMS LabWare
• Compliance with USP <643>, USP <645>, Ph. Eur. 2.2.44
Critical parameters:
• TOC: ≤ 500 ppb
• Conductivity: ≤ 1.3 µS/cm (at 25°C)
• Standard Response (SST): 85-115%
• Blank: ≤ 100 ppb
💡 Usage tips:
1. Sampling: Use dedicated TOC taps, flushing the system with water for 2-5 minutes beforehand.
2. Bubbles: Avoid air bubbles in the instrument cell, as CO₂ from air affects conductivity and TOC.
3. Calibration: Calibrate the instrument with sucrose standard (500 ppb C) regularly.
4. Interpretation: If TOC is normal but conductivity is high — problem with ion exchange resins or RO. If vice versa — problem with biofilm or organics.
5. Storage: TOC samples cannot be stored long; analysis should be performed immediately or within 24 hours under refrigeration.
⚠️ Note: USP allows testing in three stages (Stage 1, 2, 3). This utility focuses on Stage 1 (online or lab analysis without complex temperature compensation tables), which is the most common scenario for rapid control.input.csv
BatchNumber,WaterType,TOC_ppb,Conductivity_uS_cm,Temperature_C,ResponseStandard_ppb,BlankResponse_ppb WFI-2026-LOOP-001,WFI,120.0,0.9,25.0,480.0,50.0 PW-2026-TANK-002,PW,450.0,1.1,24.5,490.0,80.0 WFI-2026-FAIL-003,WFI,600.0,0.8,25.0,485.0,45.0
URS & FS — TOC Water
This document describes the user requirements and functional specification for TOCWaterQualityChecker. The utility is intended for QC laboratory use as a rule-based check of data prepared from an instrument, LIMS, ELN, MES or an approved input.csv.
URS — User Requirements Specification
| ID | Requirement | Criticality | Acceptance criterion |
|---|---|---|---|
| URS-001 | The utility shall accept input.csv with approved columns: BatchNumber, WaterType, TOC_ppb, Conductivity_uS_cm, Temperature_C, ResponseStandard_ppb, BlankResponse_ppb. | High | The CSV file is processed without manual header editing. |
| URS-002 | The utility shall perform deterministic evaluation for “TOC Water” without machine learning. | High | The same input data produce the same JSON result. |
| URS-003 | The utility shall validate mandatory fields, numeric formats, flags, ranges and domain plausibility. | High | Schema and conversion errors are explicitly reported. |
| URS-004 | The utility shall generate output.json with PASS / WARNING / FAIL statuses, source values, warnings and failures. | High | The JSON result is suitable for review, deviation investigation and integration. |
| URS-005 | The documentation shall support IQ/OQ/PQ or CSV/CSA verification. | Medium | URS/FS, CSV/JSON contract and test scenarios are supplied with the utility. |
input.csv contract
| # | Field | Sample | Purpose |
|---|---|---|---|
| 1 | BatchNumber | WFI-2026-LOOP-001 | Input parameter for deterministic QC evaluation. |
| 2 | WaterType | WFI | Input parameter for deterministic QC evaluation. |
| 3 | TOC_ppb | 120.0 | Input parameter for deterministic QC evaluation. |
| 4 | Conductivity_uS_cm | 0.9 | Input parameter for deterministic QC evaluation. |
| 5 | Temperature_C | 25.0 | Input parameter for deterministic QC evaluation. |
| 6 | ResponseStandard_ppb | 480.0 | Input parameter for deterministic QC evaluation. |
| 7 | BlankResponse_ppb | 50.0 | Input parameter for deterministic QC evaluation. |
FS — Functional Specification
| ID | Function | Implementation |
|---|---|---|
| FS-001 | CSV import | Read input.csv; validate header, column count and encoding. |
| FS-002 | Schema validation | Check required fields and permitted input values. |
| FS-003 | Rule engine | Apply domain rules, limits and system suitability/specification checks described in the source utility description. |
| FS-004 | Status aggregation | Produce final status: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance. |
| FS-005 | JSON export | Write machine-readable output.json for LIMS/ELN/MES and QA/QC review. |
OQ/PQ scenarios
- OQ-001: a valid sample row shall be processed without schema error.
- OQ-002: a missing mandatory column shall produce a schema error.
- OQ-003: a non-numeric value in a numeric field shall produce a conversion error.
- OQ-004: a critical parameter outside the limit shall produce
FAILor a critical finding. - PQ-001: user real batches shall be checked with retention of
input.csv,output.json, utility version and checksum.
Included in packages
Biopharmaceuticals Extended QC Suite
Extended QC utility coverage for mAbs, biosimilars, proteins, insulins, vaccines, mRNA/LNP, HCP, sterile release, microbiology, water, cleanroom and stability workflows.
OpenDental & Dental Materials QC Suite
QC utility package for dental medicinal products, oral-care products and dental materials: anaesthetics, antiseptics, analgesic/anti-infective products, polymers/minerals, resins, bonding workflows, microbiology, stability and material QC.
OpenLabEx QC Suite
LabEx package for laboratory QC methods: chromatography, spectroscopy, microbiology, sterility, endotoxins, particles, dissolution, water, cleanrooms and sample preparation.
Open