Syringeability Force Tester

dermal fillers endotoxins hyaluronic acid injectables particles rheology sterility
Open selection
Syringeability Force Tester — Filler Extrusion Force Test (ISO 13485)

ℹ️  The utility checks peak and average force, smoothness (CV), ergonomics.
ℹ️  Does not use machine learning — mechanical calculations only.
⚠️  CRITICAL! Force must match needle gauge. CV < 15% (smoothness).

Usage:
  SyringeabilityForceTester.exe                            → demo mode
  SyringeabilityForceTester.exe input.csv output.json      → evaluate data

Input format:
BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type

Example:
SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall

— WHY IS THIS NEEDED?
Extrusion force determines physician comfort and injection accuracy:
• Peak Force (Max Force) — force to break static friction ("breakthrough"). Too high causes needle jump and tissue trauma.
• Average Force (Avg Force) — force to maintain flow. High force causes hand fatigue during large volume treatments.
• Smoothness (CV%) — coefficient of variation of force. High CV means "stick-slip" effect (jerky motion), making smooth lines/boluses impossible.
• Needle dependence: force increases sharply with decreasing gauge (27G -> 30G) and needle length.
⚠️ CRITICAL:
• Force ≤ Specification for given gauge (e.g., ≤15N for 27G, ≤25N for 30G).
• CV ≤ 15% — guarantee of smooth extrusion without jerks.
• Ergonomic Rating: assesses combination of force and smoothness (Excellent/Good/Poor).
• Rheology Correlation: high viscosity/G' should yield predictably high force. Discrepancies may indicate needle clogging or syringe defect.
• Sterility and Endotoxins — mandatory.
⚠️ Note: Utility accounts for needle type (ThinWall has lower resistance than Regular). Smoothness analysis (CV) often more important than absolute force, as jerks during facial injection are unacceptable. Test performed on texture analyzer or universal testing machine.

input.csv

BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type
SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: SyringeabilityForceTesterAPI / object: Syringeability Force TesterCategory: Dermal fillers

1. Scope

The Syringeability Force Tester utility is used for Syringeability Force Tester. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Dermal fillers

Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility

2. Execution modes

SyringeabilityForceTester.exe                            → demo mode
SyringeabilityForceTester.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • microbiology and pathogens
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ CRITICAL! Force must match needle gauge. CV < 15% (smoothness).
  • Needle dependence: force increases sharply with decreasing gauge (27G -> 30G) and needle length.
  • ⚠️ CRITICAL:
  • Force ≤ Specification for given gauge (e.g., ≤15N for 27G, ≤25N for 30G).
  • CV ≤ 15% — guarantee of smooth extrusion without jerks.
  • Ergonomic Rating: assesses combination of force and smoothness (Excellent/Good/Poor).
  • Rheology Correlation: high viscosity/G' should yield predictably high force. Discrepancies may indicate needle clogging or syringe defect.
  • Sterility and Endotoxins — mandatory.
  • ⚠️ Note: Utility accounts for needle type (ThinWall has lower resistance than Regular). Smoothness analysis (CV) often more important than absolute force, as jerks during facial injection are unacceptable. Test performed on texture analyzer or universal testing machine.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Syringeability Force Tester with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: microbiology and pathogens, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberSYR-HA-2026-001mandatory field; used for batch/lot traceability
2Max_Force_Ndecimalas specifiedMax Force N12.5mandatory numeric value; rule comparison is performed by the utility
3Average_Force_Ndecimalas specifiedAverage Force N9.0mandatory numeric value; rule comparison is performed by the utility
4Syringe_Barrel_Diameter_mmdecimalas specifiedSyringe Barrel Diameter mm9.5mandatory numeric value; rule comparison is performed by the utility
5Needle_Gaugedecimalas specifiedNeedle Gauge27.0mandatory numeric value; rule comparison is performed by the utility
6Needle_Length_mmdecimalas specifiedNeedle Length mm13.0mandatory numeric value; rule comparison is performed by the utility
7Plunger_Speed_mm_per_mindecimalminPlunger Speed mm per min30.0mandatory numeric value; rule comparison is performed by the utility
8Temperature_Cdecimal°CTemperature C25.0mandatory numeric value; rule comparison is performed by the utility
9Viscosity_Pa_sdecimalas specifiedViscosity Pa s25.0mandatory numeric value; rule comparison is performed by the utility
10G_Prime_Padecimalas specifiedG Prime Pa450.0mandatory numeric value; rule comparison is performed by the utility
11Max_Allowable_Force_Ndecimalas specifiedMax Allowable Force N15.0mandatory numeric value; rule comparison is performed by the utility
12Force_Variation_CV_Percentdecimal%Force Variation CV %8.5mandatory numeric value; rule comparison is performed by the utility
13SterilityTestboolean / flag0/1Sterility Test0valid flag required; prohibited organisms and growth are normally expected as 0 / absent
14Endotoxin_EU_mldecimalas specifiedEndotoxin EU ml0.1numeric value; critical safety attribute compared with endotoxin limit
15Needle_Typetextas specifiedNeedle TypeThinWallmandatory value; format and acceptability are checked by the utility

CSV example

BatchNumber,Max_Force_N,Average_Force_N,Syringe_Barrel_Diameter_mm,Needle_Gauge,Needle_Length_mm,Plunger_Speed_mm_per_min,Temperature_C,Viscosity_Pa_s,G_Prime_Pa,Max_Allowable_Force_N,Force_Variation_CV_Percent,SterilityTest,Endotoxin_EU_ml,Needle_Type
SYR-HA-2026-001,12.5,9.0,9.5,27.0,13.0,30.0,25.0,25.0,450.0,15.0,8.5,0,0.1,ThinWall

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable SyringeabilityForceTester.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Syringeability Force Tester, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "SyringeabilityForceTester",
  "api": "Syringeability Force Tester",
  "batchNumber": "SYR-HA-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "SYR-HA-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Max_Force_N",
      "value": "12.5",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Average_Force_N",
      "value": "9.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Syringe_Barrel_Diameter_mm",
      "value": "9.5",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Needle_Gauge",
      "value": "27.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Needle_Length_mm",
      "value": "13.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Plunger_Speed_mm_per_min",
      "value": "30.0",
      "unit": "min",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Temperature_C",
      "value": "25.0",
      "unit": "°C",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-005Positive pathogen/microbiological flag or microbiological limit excursionCritical FAIL is produced with the offending parameter.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Dermal Fillers QC Suite

Package for injectable implants and dermal fillers: hyaluronic acid, CaHA, PLLA, PMMA, collagen and related tests.

Open