PolyLLacticAcidQualityChecker

Poly L Lactic Acid

dermal fillers endotoxins hyaluronic acid injectables particles rheology sterility
Open selection

PLLA Filler Quality Checker — Poly-L-Lactic Acid Fillers (Sculptra type) (ISO 13485, EU MDR)

ℹ️ The utility checks PLLA content, particle size (critical!), suspension time, residual monomer, polymer MW, sterility.
ℹ️ Does not use machine learning — rule-based logic only.
⚠️ CRITICAL! Particles > 150 µm granuloma risk. Suspension time ≤ 30 sec. Residual monomer ≤ 0.5%!

Usage:
PolyLLacticAcidQualityChecker.exe → demo mode
PolyLLacticAcidQualityChecker.exe input.csv output.json → evaluate data

Input format:
BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score

Example:
PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0

— WHY IS THIS NEEDED?
Poly-L-Lactic Acid (PLLA) fillers — long-acting collagen biostimulators:
• Sculptra, AestheFill, Gana etc.
• Lyophilized powder requiring reconstitution before injection
• Act by stimulating fibroblasts (non-filling effect)
• Particles must be microporous and specific size for proper integration
⚠️ CRITICAL:
• Particle Size D50 20–50 µm — optimal for phagocytosis and stimulation without migration
• Max Particle Size ≤ 150 µm — larger particles cause granulomas and subcutaneous nodules
• Suspension Time ≤ 30 sec — indicates quality of lyophilization and surface treatment (fast wetting)
• Residual Monomer (Lactic Acid) ≤ 0.5% — causes irritation and pain on injection
• Molecular Weight 10–50 kDa — affects degradation rate and duration of effect
• Sterility and Endotoxins — MANDATORY for reconstitutable injectable forms
• Lyophilizate Moisture ≤ 2% — ensures stability during storage
⚠️ Note: PLLA fillers sold as lyophilized powder. "Cake" quality (density, color, integrity) important for ease of reconstitution. PLLA particles must be strictly calibrated: too small degrade fast, too large cause inflammation. Suspension time critical for clinical practice (doctor shouldn't wait minutes to draw product). Residual monomer controlled by GC or titration.

input.csv

BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score
PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0

URS & FS — user requirements and functional specification

This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.

Utility: PolyLLacticAcidQualityCheckerAPI / object: Poly L Lactic AcidCategory: Dermal fillers

1. Scope

The Poly L Lactic Acid Quality Checker utility is used for Poly L Lactic Acid. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.

The utility does not use machine learning; decisions are produced by deterministic rules.

Categories: Dermal fillers

Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility

2. Execution modes

PolyLLacticAcidQualityChecker.exe                            → demo mode
PolyLLacticAcidQualityChecker.exe input.csv output.json      → evaluate data

3. Key controlled areas

  • assay / content
  • microbiology and pathogens
  • pH and physicochemical parameters

4. Domain limits and critical parameters

  • ⚠️ CRITICAL! Particles > 150 µm granuloma risk. Suspension time ≤ 30 sec. Residual monomer ≤ 0.5%!
  • ⚠️ CRITICAL:
  • Particle Size D50 20–50 µm — optimal for phagocytosis and stimulation without migration
  • Max Particle Size ≤ 150 µm — larger particles cause granulomas and subcutaneous nodules
  • Suspension Time ≤ 30 sec — indicates quality of lyophilization and surface treatment (fast wetting)
  • Residual Monomer (Lactic Acid) ≤ 0.5% — causes irritation and pain on injection
  • Molecular Weight 10–50 kDa — affects degradation rate and duration of effect
  • Sterility and Endotoxins — MANDATORY for reconstitutable injectable forms
  • Lyophilizate Moisture ≤ 2% — ensures stability during storage
  • ⚠️ Note: PLLA fillers sold as lyophilized powder. "Cake" quality (density, color, integrity) important for ease of reconstitution. PLLA particles must be strictly calibrated: too small degrade fast, too large cause inflammation. Suspension time critical for clinical practice (doctor shouldn't wait minutes to draw product). Residual monomer controlled by GC or titration.
Limits stated in the description must be verified against the current approved specification, pharmacopoeial monograph and registration dossier before production use.

5. URS — user requirements

IDRequirementCriticalityAcceptance criterion
URS-001The system shall accept an input.csv file for Poly L Lactic Acid with the exact columns listed in the “Input data contract” section.HighA file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error.
URS-002The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data.MediumBoth execution scenarios produce a predictable result or clear diagnostic error.
URS-003The system shall perform rule-based controls for: assay / content, microbiology and pathogens, pH and physicochemical parameters.HighEach controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML.
URS-004The system shall preserve traceability between batch/lot, source values, applied rules and final verdict.HighOutput includes batch identifier, source values, parameter statuses and critical findings.
URS-005The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review.HighJSON contains overall status, check array, warnings, failures and source-file reference.
URS-006The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks.HighThe document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval.
URS-007The system shall clearly separate technical data errors from specification nonconformities.MediumSchema/type errors are not mixed with pharmacopoeial deviations and are reported separately.

6. input.csv data contract

The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.

#ColumnTypeUnitDescriptionSampleControl rule
1BatchNumberidentifieras specifiedBatch NumberPLLA-SCULP-2026-001mandatory field; used for batch/lot traceability
2PLLA_Content_mgdecimalas specifiedPLLA Content mg150.0numeric value; compared with assay/purity limit from specification or method
3Particle_Size_D50_umdecimalµmParticle Size D50 um35.0mandatory numeric value; rule comparison is performed by the utility
4Particle_Size_D90_umdecimalµmParticle Size D90 um60.0mandatory numeric value; rule comparison is performed by the utility
5Particle_Size_Max_umdecimalµmParticle Size Max um90.0mandatory numeric value; rule comparison is performed by the utility
6Suspension_Time_secdecimalas specifiedSuspension Time sec15.0mandatory numeric value; rule comparison is performed by the utility
7Residual_Monomer_Percentdecimal%Residual Monomer %0.2mandatory numeric value; rule comparison is performed by the utility
8Molecular_Weight_Dadecimalas specifiedMolecular Weight Da25000mandatory numeric value; rule comparison is performed by the utility
9pH_ReconstituteddecimalpHp H Reconstituted7.0mandatory numeric value; rule comparison is performed by the utility
10Endotoxin_EU_mldecimalas specifiedEndotoxin EU ml0.1numeric value; critical safety attribute compared with endotoxin limit
11SterilityTestboolean / flag0/1Sterility Test0valid flag required; prohibited organisms and growth are normally expected as 0 / absent
12PackagingIntegrityTestdecimalas specifiedPackaging Integrity Test0mandatory numeric value; rule comparison is performed by the utility
13Moisture_Percentdecimal%Moisture %1.2mandatory numeric value; rule comparison is performed by the utility
14Cake_Appearance_Scoredecimalas specifiedCake Appearance Score5.0mandatory numeric value; rule comparison is performed by the utility

CSV example

BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score
PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0

7. FS — functional specification

IDFunctionImplementation description
FS-001CLI entry pointThe executable PolyLLacticAcidQualityChecker.exe supports demo mode and input.csv output.json processing mode.
FS-002CSV parserThe import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator.
FS-003Field conversionEach column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag.
FS-004Domain rule engineFor Poly L Lactic Acid, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check.
FS-005Criticality handlingCritical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS.
FS-006JSON writeroutput.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages.
FS-007Integration contractThe CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service.
FS-008Error handlingSchema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS.

8. output.json contract

The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.

{
  "utility": "PolyLLacticAcidQualityChecker",
  "api": "Poly L Lactic Acid",
  "batchNumber": "PLLA-SCULP-2026-001",
  "overallStatus": "PASS|WARNING|FAIL",
  "checkedAtUtc": "2026-05-18T00:00:00Z",
  "checks": [
    {
      "parameter": "BatchNumber",
      "value": "PLLA-SCULP-2026-001",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "PLLA_Content_mg",
      "value": "150.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Particle_Size_D50_um",
      "value": "35.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Particle_Size_D90_um",
      "value": "60.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Particle_Size_Max_um",
      "value": "90.0",
      "unit": "µm",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Suspension_Time_sec",
      "value": "15.0",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Residual_Monomer_Percent",
      "value": "0.2",
      "unit": "%",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    },
    {
      "parameter": "Molecular_Weight_Da",
      "value": "25000",
      "unit": "as specified",
      "status": "PASS|WARNING|FAIL",
      "message": "Rule-based check result"
    }
  ],
  "criticalFindings": [],
  "sourceFile": "input.csv"
}

9. OQ/PQ test scenarios

IDScenarioExpected result
TC-001Valid CSV with expected header and sample rowAll rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail.
TC-002A mandatory input.csv column is missingImport is rejected or the row receives FAIL with schema reference.
TC-003A numeric field contains text or a blank valueType conversion error is recorded; the result is not hidden as PASS.
TC-004A parameter is outside specification or critical limitCritical parameters produce FAIL; non-critical deviations produce WARNING according to the rule.
TC-005Positive pathogen/microbiological flag or microbiological limit excursionCritical FAIL is produced with the offending parameter.

10. QA/QC, CSV and change control

  • Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
  • Column names must not be changed without updating the validator and test scenarios.
  • The source CSV, output.json and utility version should be stored together as an evidence package.
  • Any change in control rules must go through change control and repeated OQ scenario verification.

Included in packages

Dermal Fillers QC Suite

Package for injectable implants and dermal fillers: hyaluronic acid, CaHA, PLLA, PMMA, collagen and related tests.