PolyLLacticAcidQualityChecker
Poly L Lactic Acid
PLLA Filler Quality Checker — Poly-L-Lactic Acid Fillers (Sculptra type) (ISO 13485, EU MDR)
ℹ️ The utility checks PLLA content, particle size (critical!), suspension time, residual monomer, polymer MW, sterility.
ℹ️ Does not use machine learning — rule-based logic only.
⚠️ CRITICAL! Particles > 150 µm granuloma risk. Suspension time ≤ 30 sec. Residual monomer ≤ 0.5%!
Usage:
PolyLLacticAcidQualityChecker.exe → demo mode
PolyLLacticAcidQualityChecker.exe input.csv output.json → evaluate data
Input format:
BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score
Example:
PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0
— WHY IS THIS NEEDED?
Poly-L-Lactic Acid (PLLA) fillers — long-acting collagen biostimulators:
• Sculptra, AestheFill, Gana etc.
• Lyophilized powder requiring reconstitution before injection
• Act by stimulating fibroblasts (non-filling effect)
• Particles must be microporous and specific size for proper integration
⚠️ CRITICAL:
• Particle Size D50 20–50 µm — optimal for phagocytosis and stimulation without migration
• Max Particle Size ≤ 150 µm — larger particles cause granulomas and subcutaneous nodules
• Suspension Time ≤ 30 sec — indicates quality of lyophilization and surface treatment (fast wetting)
• Residual Monomer (Lactic Acid) ≤ 0.5% — causes irritation and pain on injection
• Molecular Weight 10–50 kDa — affects degradation rate and duration of effect
• Sterility and Endotoxins — MANDATORY for reconstitutable injectable forms
• Lyophilizate Moisture ≤ 2% — ensures stability during storage
⚠️ Note: PLLA fillers sold as lyophilized powder. "Cake" quality (density, color, integrity) important for ease of reconstitution. PLLA particles must be strictly calibrated: too small degrade fast, too large cause inflammation. Suspension time critical for clinical practice (doctor shouldn't wait minutes to draw product). Residual monomer controlled by GC or titration.
input.csv
BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0
URS & FS — user requirements and functional specification
This document defines user requirements and functional specification for the quality-control utility. It supports deployment discussion, IQ/OQ preparation and QC workflow integration.
1. Scope
The Poly L Lactic Acid Quality Checker utility is used for Poly L Lactic Acid. The actual input.csv is the source of truth for the input structure; control criteria are defined by the executable and the domain description.
The utility does not use machine learning; decisions are produced by deterministic rules.
Categories: Dermal fillers
Tags: dermal fillers, endotoxins, hyaluronic acid, injectables, particles, rheology, sterility
2. Execution modes
PolyLLacticAcidQualityChecker.exe → demo mode PolyLLacticAcidQualityChecker.exe input.csv output.json → evaluate data
3. Key controlled areas
- assay / content
- microbiology and pathogens
- pH and physicochemical parameters
4. Domain limits and critical parameters
- ⚠️ CRITICAL! Particles > 150 µm granuloma risk. Suspension time ≤ 30 sec. Residual monomer ≤ 0.5%!
- ⚠️ CRITICAL:
- Particle Size D50 20–50 µm — optimal for phagocytosis and stimulation without migration
- Max Particle Size ≤ 150 µm — larger particles cause granulomas and subcutaneous nodules
- Suspension Time ≤ 30 sec — indicates quality of lyophilization and surface treatment (fast wetting)
- Residual Monomer (Lactic Acid) ≤ 0.5% — causes irritation and pain on injection
- Molecular Weight 10–50 kDa — affects degradation rate and duration of effect
- Sterility and Endotoxins — MANDATORY for reconstitutable injectable forms
- Lyophilizate Moisture ≤ 2% — ensures stability during storage
- ⚠️ Note: PLLA fillers sold as lyophilized powder. "Cake" quality (density, color, integrity) important for ease of reconstitution. PLLA particles must be strictly calibrated: too small degrade fast, too large cause inflammation. Suspension time critical for clinical practice (doctor shouldn't wait minutes to draw product). Residual monomer controlled by GC or titration.
5. URS — user requirements
| ID | Requirement | Criticality | Acceptance criterion |
|---|---|---|---|
| URS-001 | The system shall accept an input.csv file for Poly L Lactic Acid with the exact columns listed in the “Input data contract” section. | High | A file with the correct header is processed without manual editing; missing mandatory columns produce FAIL/import error. |
| URS-002 | The system shall support execution without arguments in demo mode and execution with input.csv output.json for user data. | Medium | Both execution scenarios produce a predictable result or clear diagnostic error. |
| URS-003 | The system shall perform rule-based controls for: assay / content, microbiology and pathogens, pH and physicochemical parameters. | High | Each controlled parameter receives a status and message; the result does not depend on hidden Excel formulas or ML. |
| URS-004 | The system shall preserve traceability between batch/lot, source values, applied rules and final verdict. | High | Output includes batch identifier, source values, parameter statuses and critical findings. |
| URS-005 | The system shall generate machine-readable output.json for LIMS/ELN/MES, batch record and QA/QC review. | High | JSON contains overall status, check array, warnings, failures and source-file reference. |
| URS-006 | The system shall support use in the client validation package: URS/FS, IQ/OQ preparation, installation and operational scenario checks. | High | The document, test scenarios and reproducible CSV/JSON flow are suitable for audit and internal approval. |
| URS-007 | The system shall clearly separate technical data errors from specification nonconformities. | Medium | Schema/type errors are not mixed with pharmacopoeial deviations and are reported separately. |
6. input.csv data contract
The source of truth for the input schema is the actual input.csv header. Column names are technical identifiers and are not translated.
| # | Column | Type | Unit | Description | Sample | Control rule |
|---|---|---|---|---|---|---|
| 1 | BatchNumber | identifier | as specified | Batch Number | PLLA-SCULP-2026-001 | mandatory field; used for batch/lot traceability |
| 2 | PLLA_Content_mg | decimal | as specified | PLLA Content mg | 150.0 | numeric value; compared with assay/purity limit from specification or method |
| 3 | Particle_Size_D50_um | decimal | µm | Particle Size D50 um | 35.0 | mandatory numeric value; rule comparison is performed by the utility |
| 4 | Particle_Size_D90_um | decimal | µm | Particle Size D90 um | 60.0 | mandatory numeric value; rule comparison is performed by the utility |
| 5 | Particle_Size_Max_um | decimal | µm | Particle Size Max um | 90.0 | mandatory numeric value; rule comparison is performed by the utility |
| 6 | Suspension_Time_sec | decimal | as specified | Suspension Time sec | 15.0 | mandatory numeric value; rule comparison is performed by the utility |
| 7 | Residual_Monomer_Percent | decimal | % | Residual Monomer % | 0.2 | mandatory numeric value; rule comparison is performed by the utility |
| 8 | Molecular_Weight_Da | decimal | as specified | Molecular Weight Da | 25000 | mandatory numeric value; rule comparison is performed by the utility |
| 9 | pH_Reconstituted | decimal | pH | p H Reconstituted | 7.0 | mandatory numeric value; rule comparison is performed by the utility |
| 10 | Endotoxin_EU_ml | decimal | as specified | Endotoxin EU ml | 0.1 | numeric value; critical safety attribute compared with endotoxin limit |
| 11 | SterilityTest | boolean / flag | 0/1 | Sterility Test | 0 | valid flag required; prohibited organisms and growth are normally expected as 0 / absent |
| 12 | PackagingIntegrityTest | decimal | as specified | Packaging Integrity Test | 0 | mandatory numeric value; rule comparison is performed by the utility |
| 13 | Moisture_Percent | decimal | % | Moisture % | 1.2 | mandatory numeric value; rule comparison is performed by the utility |
| 14 | Cake_Appearance_Score | decimal | as specified | Cake Appearance Score | 5.0 | mandatory numeric value; rule comparison is performed by the utility |
CSV example
BatchNumber,PLLA_Content_mg,Particle_Size_D50_um,Particle_Size_D90_um,Particle_Size_Max_um,Suspension_Time_sec,Residual_Monomer_Percent,Molecular_Weight_Da,pH_Reconstituted,Endotoxin_EU_ml,SterilityTest,PackagingIntegrityTest,Moisture_Percent,Cake_Appearance_Score PLLA-SCULP-2026-001,150.0,35.0,60.0,90.0,15.0,0.2,25000,7.0,0.1,0,0,1.2,5.0
7. FS — functional specification
| ID | Function | Implementation description |
|---|---|---|
| FS-001 | CLI entry point | The executable PolyLLacticAcidQualityChecker.exe supports demo mode and input.csv output.json processing mode. |
| FS-002 | CSV parser | The import module reads CSV, validates header, column presence/order, value count and encoding. Decimal values are expected with a dot separator. |
| FS-003 | Field conversion | Each column is converted to the expected type: identifier, text, decimal number, date/time or boolean flag. |
| FS-004 | Domain rule engine | For Poly L Lactic Acid, explicit rules are applied: range, minimum, maximum, absence of prohibited flag, data completeness or calculation-based check. |
| FS-005 | Criticality handling | Critical violations produce FAIL; non-critical deviations and incomplete data produce WARNING; full conformance produces PASS. |
| FS-006 | JSON writer | output.json stores overall status, per-parameter results, source values, warnings, failures and diagnostic messages. |
| FS-007 | Integration contract | The CSV → JSON format is stable for invocation from LIMS/ELN/MES, scheduled task or wrapper service. |
| FS-008 | Error handling | Schema error, missing file, non-numeric value or JSON write failure returns diagnosable error without silent PASS. |
8. output.json contract
The output file must be suitable for automated processing, audit review and correlation with the source input.csv row.
{
"utility": "PolyLLacticAcidQualityChecker",
"api": "Poly L Lactic Acid",
"batchNumber": "PLLA-SCULP-2026-001",
"overallStatus": "PASS|WARNING|FAIL",
"checkedAtUtc": "2026-05-18T00:00:00Z",
"checks": [
{
"parameter": "BatchNumber",
"value": "PLLA-SCULP-2026-001",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "PLLA_Content_mg",
"value": "150.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Particle_Size_D50_um",
"value": "35.0",
"unit": "µm",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Particle_Size_D90_um",
"value": "60.0",
"unit": "µm",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Particle_Size_Max_um",
"value": "90.0",
"unit": "µm",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Suspension_Time_sec",
"value": "15.0",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Residual_Monomer_Percent",
"value": "0.2",
"unit": "%",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
},
{
"parameter": "Molecular_Weight_Da",
"value": "25000",
"unit": "as specified",
"status": "PASS|WARNING|FAIL",
"message": "Rule-based check result"
}
],
"criticalFindings": [],
"sourceFile": "input.csv"
}9. OQ/PQ test scenarios
| ID | Scenario | Expected result |
|---|---|---|
| TC-001 | Valid CSV with expected header and sample row | All rows are processed; output contains PASS/WARNING/FAIL and parameter-level detail. |
| TC-002 | A mandatory input.csv column is missing | Import is rejected or the row receives FAIL with schema reference. |
| TC-003 | A numeric field contains text or a blank value | Type conversion error is recorded; the result is not hidden as PASS. |
| TC-004 | A parameter is outside specification or critical limit | Critical parameters produce FAIL; non-critical deviations produce WARNING according to the rule. |
| TC-005 | Positive pathogen/microbiological flag or microbiological limit excursion | Critical FAIL is produced with the offending parameter. |
10. QA/QC, CSV and change control
- Before production use, the client records executable version, checksum, specification/monograph version, test CSV, expected JSON and IQ/OQ results.
- Column names must not be changed without updating the validator and test scenarios.
- The source CSV, output.json and utility version should be stored together as an evidence package.
- Any change in control rules must go through change control and repeated OQ scenario verification.
Included in packages
Dermal Fillers QC Suite
Package for injectable implants and dermal fillers: hyaluronic acid, CaHA, PLLA, PMMA, collagen and related tests.
Open