AutomatedLiquidHandlingQCQualityChecker

Automated Liquid Handling QC

LabEx laboratory QC CSV→JSON URS & FS chromatography elemental impurities utilities / cleanroom
Open selection
Automated Liquid Handling QC Quality Checker — Quality Control for Automated Liquid Handling Systems
ℹ️ Utility checks critical parameters of automated liquid handlers (Hamilton, Tecan, Eppendorf, etc.):
• Accuracy (Systematic Error) — deviation of mean volume from target
• Precision (CV / Random Error) — reproducibility of dispensing
• Compliance with ISO 8655 and USP <1058> (instrument qualification)
⚠️ CRITICAL: Dispensing errors >3% (accuracy) or CV >1% (precision) can lead to incorrect results in HPLC, ICP-MS, ELISA, and other assays!
Usage:
AutomatedLiquidHandlingQCQualityChecker.exe → demo mode (console output)
AutomatedLiquidHandlingQCQualityChecker.exe input.csv output.json → evaluate your data
Input format:
BatchNumber,TargetVolumeUL,MeanMeasuredVolumeUL,StandardDeviationUL,NumberOfReplicates,WaterDensityMGPerUL
Example:
CAL-HAMILTON-2026-001,100.0,99.8,0.4,10,0.9982
Supported equipment types:
• Automated Workstations (Hamilton STAR, Tecan Fluent, Beckman Biomek)
• Liquid Handlers (Eppendorf epMotion, Gilson Prep)
• Calibration Events (Regular gravimetric calibration)
— WHY IS THIS NEEDED?
Automated dispensing is the foundation of reproducibility in modern pharma labs:
• Sample preparation for elemental analysis (ICP-MS) requires high dilution accuracy
• Microplate setup for ELISA/biological tests requires low variability (CV)
• USP <1058> requires regular qualification (PQ) of all measurement systems, including pipettes and handlers
⚠️ CRITICAL:
• Accuracy: Deviation of mean volume from target. Limit typically ±3% (volume-dependent).
• Precision (CV): Coefficient of variation. Limit typically <1% for volumes >50 µL.
• Gravimetric method: Primary verification method. Uses water density (Z-factor) to convert mass to volume.
• Temperature and humidity: Must be controlled during calibration (ISO 8655).
Key features:
• Automatic calculation of % Error (Accuracy) and Coefficient of Variation (CV)
• Comparison with configurable pharmacopoeial/internal limits
• Support for gravimetric data (mass -> volume via density)
• LIMS-compatible report generation (LabWare) for qualification status (Pass/Fail)
Critical parameters:
• Accuracy: Within ±3.0% (example for 100 µL)
• Precision (CV): ≤ 1.0% (example for 100 µL)
• Water Density: Temperature-corrected (e.g., 0.9982 mg/µL at 20°C)
💡 Usage tips:
Perform gravimetric calibration at least quarterly or after maintenance.
Use analytical balances with appropriate weighing range (minimum weight).
For small volumes (<10 µL), use evaporation traps and specialized techniques.
If CV is high — check tips, seals, and air bubbles.
If Accuracy is low — perform software calibration (calibration factor) of the handler.
⚠️ Note: This utility does not control the robot directly but analyzes gravimetric test data loaded by the operator or exported from balance software. It is a key element of the laboratory's Quality Assurance (QA) system.

input.csv

BatchNumber,TargetVolumeUL,MeanMeasuredVolumeUL,StandardDeviationUL,NumberOfReplicates,WaterDensityMGPerUL
CAL-HAMILTON-2026-001,100.0,99.8,0.4,10,0.9982
CAL-TECAN-2026-002,50.0,49.5,0.8,10,0.9982
CAL-EPPENDORF-2026-003,200.0,200.5,0.5,10,0.9982

URS & FS — Automated Liquid Handling QC

This document describes the user requirements and functional specification for AutomatedLiquidHandlingQCQualityChecker. The utility is intended for QC laboratory use as a rule-based check of data prepared from an instrument, LIMS, ELN, MES or an approved input.csv.

URS — User Requirements Specification

IDRequirementCriticalityAcceptance criterion
URS-001The utility shall accept input.csv with approved columns: BatchNumber, TargetVolumeUL, MeanMeasuredVolumeUL, StandardDeviationUL, NumberOfReplicates, WaterDensityMGPerUL.HighThe CSV file is processed without manual header editing.
URS-002The utility shall perform deterministic evaluation for “Automated Liquid Handling QC” without machine learning.HighThe same input data produce the same JSON result.
URS-003The utility shall validate mandatory fields, numeric formats, flags, ranges and domain plausibility.HighSchema and conversion errors are explicitly reported.
URS-004The utility shall generate output.json with PASS / WARNING / FAIL statuses, source values, warnings and failures.HighThe JSON result is suitable for review, deviation investigation and integration.
URS-005The documentation shall support IQ/OQ/PQ or CSV/CSA verification.MediumURS/FS, CSV/JSON contract and test scenarios are supplied with the utility.

input.csv contract

#FieldSamplePurpose
1BatchNumberCAL-HAMILTON-2026-001Input parameter for deterministic QC evaluation.
2TargetVolumeUL100.0Input parameter for deterministic QC evaluation.
3MeanMeasuredVolumeUL99.8Input parameter for deterministic QC evaluation.
4StandardDeviationUL0.4Input parameter for deterministic QC evaluation.
5NumberOfReplicates10Input parameter for deterministic QC evaluation.
6WaterDensityMGPerUL0.9982Input parameter for deterministic QC evaluation.

FS — Functional Specification

IDFunctionImplementation
FS-001CSV importRead input.csv; validate header, column count and encoding.
FS-002Schema validationCheck required fields and permitted input values.
FS-003Rule engineApply domain rules, limits and system suitability/specification checks described in the source utility description.
FS-004Status aggregationProduce final status: FAIL for critical failure, WARNING for non-critical deviation, PASS for conformance.
FS-005JSON exportWrite machine-readable output.json for LIMS/ELN/MES and QA/QC review.

OQ/PQ scenarios

  • OQ-001: a valid sample row shall be processed without schema error.
  • OQ-002: a missing mandatory column shall produce a schema error.
  • OQ-003: a non-numeric value in a numeric field shall produce a conversion error.
  • OQ-004: a critical parameter outside the limit shall produce FAIL or a critical finding.
  • PQ-001: user real batches shall be checked with retention of input.csv, output.json, utility version and checksum.

Included in packages

Biopharmaceuticals Extended QC Suite

Extended QC utility coverage for mAbs, biosimilars, proteins, insulins, vaccines, mRNA/LNP, HCP, sterile release, microbiology, water, cleanroom and stability workflows.

Open

LabEx QC Suite

LabEx package for laboratory QC methods: chromatography, spectroscopy, microbiology, sterility, endotoxins, particles, dissolution, water, cleanrooms and sample preparation.

Open

Radiology QC Suite

QC package for radiology and diagnostic imaging: radiopharmaceuticals, PET/SPECT, contrast media, iodinated and gadolinium products, plus particle, sterility and endotoxin checks.

Open